![Dräger Infinity Delta Instructions For Use Manual Download Page 259](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779259.webp)
A
CCURACY
VF8
D
ELTA
/D
ELTA
XL/K
APPA
15-3
Accuracy
To optimize cardiac output measurement:
Follow the recommendations made by the manufacturer. Dräger recommends
you place the prefilled syringes or the closed injectate delivery system into an
ice bath.
Check the ice bath regularly and add ice as needed to maintain a temperature
between 0
°
C and 5
°
C. Dräger recommends cold (refrigerated) saline
solution.
Verify the injectate volume.
Verify the computation constant. An incorrect computation constant is a
common cause of error.
Use an in-line injectate system. Systems that measure the temperature of the
injectate in the ice bath may introduce error, since the injectate temperature
changes in the time between its removal from the ice bath and injection. Use
an in-line temperature sensor to eliminate this source of error.
If you hand-fill your syringes, fill them with the same volume each time. The
recommended amount is 10cc for adults and 5cc for pediatric patients. Avoid
touching the body of the syringe. The warmth of your hand will warm the
injectate very quickly.
Inject the entire volume in one swift, continuous motion.
Perform the injection at end-expiration. Taking successive cardiac output
measurements at different points in the respiratory cycle can give different
measurements, especially for patients on mechanical ventilators.
Discard results that are widely different from the general trend, and those
associated with irregularly-shaped (for example, notched) curves.
NOTE:
If you use a room temperature injectate, use 10 cc for the injectate volume,
unless clinically
contraindicated. However, measurements are more reliable when using
cold (refrigerated) saline solutions.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...