IBP S
ETUP
VF8
D
ELTA
/D
ELTA
XL/K
APPA
14-17
Pressure labels are color-coded within this window to indicate their status. To assign a
label, the hemodynamic pod or Y-cable must be connected to the monitor.
Pressure Label Conflicts
Each pressure label can be assigned to one location. If you try to reuse a label, the
monitor displays a caution informing you that the label is in use and asks if you want
to continue. If you choose
YES
, the monitor puts the label in the currently selected
box and places an automatic pressure label (P1a
−
P3d) in the previous location.
Pod-Monitor Label Conflicts
Hemo2, Hemo4, PiCCO pods, and MPod – Quad Hemo and the monitor store
pressure label assignments. When a pod with previously stored labels is connected to a
monitor, the pod and the monitor may be storing different pressure labels for the same
channel, causing a conflict.
The label stored in a pod prevails if a transducer is connected to the pod. The pod
keeps the label, and the monitor assigns that parameter label to the
Pressure Labels
screen. If a transducer is not connected to the pod, the label stored in the monitor
memory has priority.
Color-Coding for Pressure Labels
Background
Text
Status
Black
Green
• Pod or Y-cable connected to monitor
• Transducer connected
White
• Pod or Y-cable connected to monitor
• Transducer not connected
Gray
White
• Pod or Y-cable not connected to monitor
• Transducer not connected
NOTE:
For Hemo2 pods, the pressure label screen shows two ghosted fields, because
the pod monitors two pressures.
NOTE:
The HemoMed pod does not store any pressure labels assigned by the
monitor.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...