RESP Setup Menu
All impedance respiration functions are controlled from the RESP setup menu, which
you can open in one of two ways:
To access the RESP setup menu
Click on the
RESP
parameter box on the main screen
or
1. Press the
Menu
fixed key on the front of the monitor.
2. Click on
Patient Setup
. A list of available patient setup functions appears.
3. Click on
Parameters
in the second column. A list of available parameters
appears.
4. Scroll to
RESP
and click to display the RESP
setup menu.
Quick Reference Table
–
Respiration Setup
The Respiration Setup Menu
Menu Item
Description
Settings
RESP Lead
Determines respiration lead
I
,
II
Mode
Determines processing mode for
breath-related impedance changes
WARNING: If the respiration
waveform size is set too low,
shallow breaths may not be
counted. If it is set too high,
cardiac artifact will be counted as
breaths. Therefore, always use
the respiration marker to verify
breath detection at the desired
amplitude.
•
Auto
– Optimal breath-detection threshold
calculated at the beginning of RESP
monitoring. Intended for patients with
regular breathing patterns.
•
Manual
– The monitor does not set a
breath-detection threshold at the
beginning of RESP monitoring. Instead,
the adjustments you make to the
waveform Size (see the
Size
setting,
below) set the monitor’s breath detection
sensitivity. Intended for adult or pediatric
patients whose breathing patterns show
excessive variation; or for neonates whose
breathing rhythms tend to be irregular, and
whose respiration signals may not be
reliably detected in
Auto
mode.
Size
Auto
mode –
Adjusts the waveform
size.
Manual
mode –
Adjusts the
waveform size and the breath-
detection threshold.
•
5 - 100 %
(in 5 % increments)
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...