Overview
Arrhythmia monitoring is available for adult and pediatric patients. The mode you
select (Full, Basic, or OFF) determines the events processed. Full arrhythmia is a
locked option which must be activated by your hospital’s technical personnel.
Arrhythmia monitoring is not available for neonates. Arrhythmia monitoring is only
available for adult and pediatric patients.
The monitor matches incoming beats against beats previously recorded and stored in a
reference template. Through this process, the monitor can verify an arrhythmia
event’s occurrence, classify it, and draw clinically useful conclusions based on the
frequency and morphology of the signal. The monitor considers all beats questionable
if a baseline shift exceeds specified limits.
The monitor uses the results of QRS processing for arrhythmia analysis. During
multiple-lead arrhythmia processing, each lead’s QRS complexes are measured and
compared against its learned dominant normal beat. The monitor classifies beats based
on information acquired from all available leads.
WARNING: Do not rely solely on ECG with seizure-prone
patients. Electrical artifacts of non-cardiac origin, such as
seizure, may prevent detection of certain arrhythmias.
NOTE:
Arrhythmia detection may not work properly in all patients. The monitor
classifies only QRS complexes > 0.20 mV, for widths > 70 ms. An artifact condition
(
ARTF
) may occur when the ECG signal does not meet these minimums. While
continuing to monitor HR, you can turn off ARR monitoring for patients whose QRS
complexes do not meet these minimums.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...