ECG Signal Processing and Display
The monitor identifies QRS complexes with amplitudes between 0.2 and 5.0 mV and
a QRS width of 70 to 120 ms for adults (or 40 to 100 ms for neonates; see note on
page 8). It calculates heart rates within a range of 15 to 300 beats per minute, using the
R-R intervals of the last 10 seconds and disregarding the two longest and two shortest
R-R intervals. It averages the remaining intervals, and displays the result as the
current heart rate in the HR parameter box on the main screen.
During dual-channel processing, the monitor assigns a weight to each channel
depending on its level of artifact (the cleaner channel receives a greater weight). When
the noise in one channel exceeds a certain level, the channel is excluded from the
composite signal, and the monitor effectively shifts to single-channel processing. The
level of artifact on an analyzed channel determines the weight that channel receives (a
cleaner channel receives a greater weight). When the noise in one channel exceeds a
certain level, the channel is excluded from the composite signal, and the monitor
effectively shifts to single-channel processing. If both channels show excessive noise,
the message
ECG Artifact
appears until at least one channel is sufficiently noise-free.
During brief artifact, the heart rate is blanked. During extended artifact, the
heart rate value is displayed as * * *.
When artifact clears, the monitor resumes QRS processing but does not
initiate a relearn.
When you enable arrhythmia monitoring (not available for neonatal patients), the HR
parameter box changes accordingly. If you select
Basic
, the four basic arrhythmia
calls ASY, VF, VT, and ARTF are available for display. If you select
Full
and you
configure the main screen to display two or more ECG channels, a separate
ARR
parameter box appears below the
HR
parameter box (see page 9 for information about
selecting an arrhythmia mode).
Basic Arrhythmia Mode
Full Arrhythmia Mode
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...