Lead color-coding for 3-, 5-, 6-lead monitoring
TruST Twelve-Lead Configuration
Overview
Infinity TruST is 12-lead ECG monitoring acquired using a MultiMed, a MultiMed
Plus, or a MultiMed OR and 6 (instead of 10) electrodes. The TruST leads are
constructed from other leads and are intended for real-time assessment of ST segment
changes. TruST is available in adult and pediatric modes, but not in neonatal mode.
In general, the signal from a measured lead provides information common to other
leads. When this information is appropriately combined, the signal of leads not
otherwise configured can be inferred. Appropriate use of this information allows ECG
signals to be derived for leads that are not physically available. This type of lead
derivation has a high degree of correlation with measured leads, but should not be
used for diagnostic 12-lead ECG analysis.
T
RU
ST
electrodes are placed in the six-lead standard configuration. As with the six-
lead pod, waveforms from eight leads can be viewed on the Delta monitor, but
T
RU
ST
also processes and displays four additional lead waveforms. These TruST leads are
viewable in the same fashion All Leads display (see Show All Leads on page 8-20) as
the conventional leads. See “Three-, Five-, Six- and TruST® Twelve-Lead
Configurations” on page 8-9 for Six-Lead Standard/Infinity TruST electrode
placement.
Lead color-coding for 3-, 5-, 6-lead monitoring
ECG Lead
IEC
AHA/US
LA
Yellow
Black
LL
Green
Red
RA
Red
White
RL
Black
Green
V
White
Brown
V+
Gray and white
Gray and brown
WARNING:
Refer to the conventional lead if the QRS morphology
of a TruST lead differs from that of its equivalent
conventional lead.
Do not select TruST leads for ECG signal processing.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...