5-14
D
ELTA
/D
ELTA
XL/K
APPA
VF8
5 A
LARMS
Alarm Limits Shortcut
Each parameter setup menu has an alarm menu selection which accesses the Alarm
Limits table, targeting associated parameters on the Alarm Limits
table. Exiting the
Alarm Limits table returns you to the parameter setup screen.
ST and Arrhythmia Alarms
ST and arrhythmia parameters have their own alarm limits configuration screens,
which you can access by selecting the
ST
or
ARR
control button at the bottom of the
alarm limits table (see page 5-6). Refer to chapters 10, and 9, for more information
about ST and arrhythmia tables.
Alarm History Table
The monitor stores up to 50 physiological alarm events for each patient. Events are
deleted when the patient is discharged. Data is stored in the monitor and remains with
the patient during Pick and Go transports. Data also survives power shutdowns. The
Alarm History table records all high-priority and medium-priority alarms, every
activation and deactivation of cardiac bypass mode, every change of patient category,
and records whenever you perform an
All Alarms Off
or
Alarm Silence
.
To access the Alarm History table
1. Press the
Fast Access
fixed key.
2. Click on
Alarm History
to display the Alarm History table.
Archive
Allows you to store and/or record automatically
an alarm event for the selected parameter.
You can later review stored alarms on the
Event Recall screen.
NOTE
: You cannot turn the Archive option off
for asystole and ventricular fibrillation.
• Store
• Record
• Str./Rec.
• OFF
Alarm Limits Table
Function
Description
Available Settings
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...