Supplement for Infinity Delta Series Software
88
Supplement – Infinity
®
Delta Series Software VF10.1
Electromagnetic Emissions
This equipment is intended for use in the electromagnetic environment specified below. The
user of this equipment should assure that is used in such an environment.
Emissions:
Compliance according to:
Electromagnetic environment:
RF emissions (CISPR 11) Group 1
The monitor only uses RF energy
internally. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment. Also
see the following note.
CISPR emissions
Classification
Delta - Class B
Delta XL - Class B
Kappa - Class A
Infinity Docking Station (IDS),
integrated MIB - Class A
SCIO Four Module - Class A
BISx pod - Class A
The equipment is suitable for use in
all establishments including
domestic establishments and those
directly connected to the public low-
voltage power supply network that
supplies buildings used for domestic
purposes.
Harmonic emissions
(IEC 61000-3-2)
Class A
Voltage fluctuations /
flicker (IEC 61000-3-3)
Complies
NOTE
When the monitor is configured with the wireless option, that option's radio emits electromagnetic
energy in order to communicate with the Infinity network. This may affect nearby equipment. See
the documentation that accompanies the wireless option's radio for further details.
CAUTION
Do not place monitor with radio operational next to paced patients.