Supplement – Infinity
®
Delta Series Software VF10.1
83
Supplement for Infinity Delta Series Software
Surface disinfection with cleaning
1
Remove soiling immediately. Use a cloth dampened with disinfectant to remove soiling.
2
Perform surface disinfection.
3
After the product has been exposed to the disinfectant for the specified contact time, remove residual
disinfectant.
4
Wipe with a cloth dampened with water (preferably drinking-water quality). Allow the product to dry.
5
Check the product for visible soiling. Repeat steps 1 to 5 if necessary.
6
Check the product for visible damage and replace if necessary.
Cleaning and Disinfecting
IBP - Infinity PiCCO Pod
Agents tested by Dräger and shown to have no harmful effect at the time of testing on the materials utilized
in the Infinity PiCCO Pod include:
– Isopropyl alcohol 40%
Higher concentrations could damage:
– Compliance™; this cleaning agent may discolor soft plastic material
– Sporox II
– Dismozon pur
WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
– Damage to the device
– Electric shock
– Device malfunctions
Ensure that no liquid penetrates the device.
CAUTION
Do not autoclave the Infinity PiCCO Pod.
CAUTION
If using alcohol, it should only be a 40% diluted solution.