Supplement – Infinity
®
Delta Series Software VF10.1
37
Supplement for Infinity Delta Series Software
Quick Reference Table - SpO
2
Setup
New Status Message
The following status message applies to any monitor using a Masimo SET pod.
Standalone Pulse Oximeter
As of the Infinity Delta Series Software VF10.1 release, is no longer supported.
Menu item
Description
Settings
Pulse Tone
Volume
Sets the volume of the
pulse tone.
NOTE
: You can also set
the pulse tone volume
from the ECG menu.
OFF
, 5, 10, 20, 30, 40, 50, 60, 70, 80, 90, or 100
SpO2 Alarm
Accesses SpO
2
alarms
on the
Alarm Limits
table.
– SpO
2
OxiSure = 20 – 100% in increments of 1%
– Pulse OxiSure = 30 – 240% in increments of 5%
– SpO
2
Masimo = 20 – 100% in increments of 1%
– Pulse Masimo = 30 – 240% in increments of 5%
– SpO
2
Nellcor = 20 – 100% in increments of 1%
– Pulse Nellcor = 30 – 240% in increments of 5%
Message
Possible Cause
Suggested Action
Replace SpO2 Sensor/Cable
The sensor or the cable is
expired or defective.
Replace the sensor or the cable.