Supplement for Infinity Delta Series Software
ECG Setup Menu
Menu selection
Description
Available Settings
The
ECG Options
submenu
This submenu allows you to configure the following functions:
Pacer Detection
Determines the monitor’s ability to
identify pacemaker pulse. Allows user
to enable/disable pacer detection or
choose more advanced
Fusion
selection.
CAUTION
:
Fusion
mode pacer
detection is not intended for use with
large-signal, unipolar pacemakers. It is
intended for use only with bipolar
pacemakers.
In basic mode:
– Select
ON
to enable pacemaker
detection.
– Select
OFF
(default) to disable
pacemaker detection.
Pacer Detection
(continued)
Observe the following:
– Select
Fusion
mode only in
situations where it becomes
necessary to suppress repeated
false asystole and/or false low heart
rate alarms.
– Before selecting
Fusion
mode, be
certain that the patient has a bipolar
pacemaker (external or implanted)
and that it is accurately
programmed as appropriate for that
patient.
– Do not use
Fusion
mode if you are
uncertain as to what type of
pacemaker is being used.
In advanced mode:
– Select
ON
to enable pacemaker
detection.
– Select
OFF
(default) to disable
pacemaker detection.
– Select
Fusion
to enable pacer
detection, but minimize pacer tail
rejection to reduce missed detection of
pseudo-fused paced beats, which
results in false asystole alarms.