Pocket Guide
Evita® V800 /
Evita® V600
Software version 1.n
D-21865-2020
Page 1: ...Pocket Guide Evita V800 Evita V600 Software version 1 n D 21865 2020 ...
Page 2: ...ute for the Instructions for Use It is for informational purposes only Any use of the device requires full understanding and strict observation of the Instruction for Use The availability of functionality depends on the country and availability Options may not be included in a present device or may come at extra costs The pocket guide may not reflect the latest software version of the device Conte...
Page 3: ...ntilator Evita is intended for the ventilation of adults adolescents children infants and neonates This device provides mandatory ventilation modes and ventilation modes for supporting spontaneous breathing as well as ventilation monitoring ...
Page 4: ...QUICK GUIDE EVITA V800 EVITA V600 04 Product description Intensive care Ventilator Display unit Ventilation unit Gas supply unit GS500 Power supply unit PS500 Trolley with 4 brakes Evita V800 ...
Page 5: ... 05 Product description Display unit front Alarm bar Touch screen Indicator for power and batteries Alarm silence 2 minutes ON OFF switch Rotary knob touch turn confirm ...
Page 6: ...QUICK GUIDE EVITA V800 EVITA V600 06 Product description Display unit rear USB ports HDMI port USB port USB ports LAN ports COM ports Swiveling of display unit ...
Page 7: ... 07 Product description Ventilation unit front Flap Expiration valve Water trap Inspiration valve Emergency air intake Operation display of ventilation Expiratory flow sensor Nebulizer port ...
Page 8: ...QUICK GUIDE EVITA V800 EVITA V600 08 Operating concept Screen general structure Header bar Monitoring area Therapy bar Main menu bar ...
Page 9: ...s dialog Trends data dialog Sensors dialog Procedures dialog System set up dialog Patient category Ventilation mode Therapy type Alarm messages Power supply charge status indicator of battery Button to open other modes Button to Start Stop ventilation Start Stop O2 therapy ...
Page 10: ...rms The following colors are used to identify the priority of alarms Colour Alarm priority Action required Red High Immediate action is necessary in order to avert an acute danger Yellow Medium Prompt action is necessary in order to avert a danger Turquoise Low Attention is necessary ...
Page 11: ... confirmation may be performed ends the background lighting flashes Blue After the device is turned on the background lighting shines for 2 seconds Touch the control element The control element turns orange Turn the rotary knob to the right or to the left to make the setting The value is displayed in the control element Confirm by pressing the rotary knob The setting is applied and the control ele...
Page 12: ...The mains power supply and the gas supply are connected The main switch is turned on 1 Press the ON OFF key on the display unit The system is started The start dialog is displayed 2 Touch the New patient button if a new patient is admitted or 3 Touch the Current patient button to continue with existing settings Ventilation unit left side Main switch 1 3 2 ...
Page 13: ... Follow the instructions in the information field and touch the button OK when asked to do so 1 3 2 4 2 1 Starting the system test 1 Touch the Start button in the System test line 2 Confirm with the rotary knob 3 Answer the questions in the information field by touching the Yes or No button After a successful system test has been performed a message appears for continuing the breathing circuit tes...
Page 14: ...imits and touch the button Start ventilation 5 Confirm with the rotary knob 1 1 3 3 5 1 Touch the button for the relevant patient category 2 Confirm with the rotary knob 3 Touch the button for height 4 Set the body height by turning the rotary knob and push to confirm 5 Touch the button for the relevant therapy type Tube NIV O2 therapy 6 Confirm with the rotary knob 7 Check the alarm limits and to...
Page 15: ...n emergency situation therapy can be started immediately without operational readiness being checked 1 Touch the button Quick start ventilation The Patient and therapy step is displayed 2 Press the rotary knob The therapy is started The main screen is displayed ...
Page 16: ...s 1 2 3 2 1 Touch the relevant tab 2 The colour of the tab turns orange if necessary adjust the ventilation parameter 3 Confirm with the rotary knob The color of the tab turns white 1 1 Touch the corresponding therapy control colour turns orange 2 Set the value by turning the rotary knob and push to confirm ...
Page 17: ...e are displayed and can be configured 2 The current settings are displayed in an information field next to the button 1 Touch the field for the therapy type in the header bar 2 Touch the button of the relevant therapy type Tube NIV O2 therapy 3 Confirm with the rotary knob The device is in standby mode 4 Start the therapy by touching the Start ventilation button 2 4 1 ...
Page 18: ...gory only the SPN CPAP and PC CMV ventilation modes can be selected Patient side accessories such as oxygen masks or nasal cannulas can be used for O2 therapy During O2 therapy only the O2 concentration the inspiratory flow and the inspiratory pressure are monitored The alarm limits for the parameters VT MVe RR Paw and Tapn are not active The alarm limits for O2 monitoring are automatically set by...
Page 19: ... the therapy bar A pop up window opens 2 Touch the button Stop ventilation 3 Confirm with the rotary knob 4 The following message is displayed in the header bar Standby mode activated 5 Touch the Alarm reset button in the header bar 6 Confirm with the rotary knob 1 2 4 3 6 5 ...
Page 20: ...ignals Display of alarms Acoustic alarm signals Alarm bar Alarm message field Blinking parameter field More alarms The primary acoustic alarm system produces different alarm signals The acoustic alarm signal can be suppressed for 2 minutes maximum 1 Press the alarm silence key 1 ...
Page 21: ...es which are no longer active 1 Open the Alarm dialog page 2 Touch the Reset all alarms button 3 Confirm with the rotary knob 1 Touch the Alarms button in the main menu bar The Alarm limits dialog page is preselected 2 Touch the corresponding button for the upper or lower alarm limit 3 Set the value by turning the rotary knob and confirm 1 1 2 ...
Page 22: ...CO2 sensor into the socket at the rear of the ventilation unit 2 Insert the cuvette into the patient port of the Y piece 3 The cuvette windows are facing to the side Position the cuvette vertically to avoid condensate accumulation 4 Place the CO2 sensor onto the cuvette The cable is facing towards the device 4 2 3 2 ...
Page 23: ... has warmed up Time until availability 10 s Time until specified accuracy is reached 120s 1 Touch the Sensors button in the main menu bar 2 Touch the CO2 sensor tab 3 Select the cuvette type used either reusable or disposable 4 Start zero calibration 5 Activate CO2 monitoring 1 4 2 5 ...
Page 24: ...ngapore 117525 Tel 65 6872 9288 Fax 65 6259 0398 asia pacific draeger com REGION NORTH AMERICA Draeger Inc 3135 Quarry Road Telford PA 18969 1042 USA Tel 1 800 4DRAGER 1 800 437 2437 Fax 1 215 723 5935 info usa draeger com REGION CENTRAL AND SOUTH AMERICA Dräger Indústria e Comércio Ltda Al Pucurui 51 Tamboré 06406 100 Barueri SP Tel 55 11 4689 4900 relacionamento draeger com Manufacturer Drägerwe...