Dräger Evita V Series Quick Start Manual Download Page 1

D

-5759-2018

Disassembly

/Assembly

 

Flow Sensor

Open the flap by lifting the lower 

edge upwards. 

Close the flap.

Push the flow sensor as far as it 

possible go to the left.

Push the flow sensor to 

 

the right into the socket.

Remove the expiratory valve 

 

from the fitting.

Turn the locking ring as far as 

possible to the left.

Remove and empty the water 

 

trap container. 

Connect the water trap container.

Remove the diaphragm.

Fit the flow sensor 

 

into the socket.

Remove the flow sensor 

 

from the socket.

Remove the sensor sleeve 

from the expiratory valve.

91 

03 

106 

19.11-1 

A

LE 

Subject 

to 

modifications 

© 

2019 

Dr

äg

er

w

er

A

C

o. 

K

G

aA

Attach the diaphragm onto the 

 

edge of the expiratory valve housing. 

 

Make sure that it fitts properly.

Attach the flow sensor sleeve 

 

on the expiratory valve.

Turn the locking ring as far 

 

as possible to the right 

 

until it clicks into place.

Turn the locking ring as far as 

possible to the left, push the 

 

expiratory valve into the fitting. 

Dräger Evita

®

 V-Series

Flow Sensor and Expiratory Valve

Disassembly/Assembly

Disassembly

/Assembly

 

Expiratory Valve*

*  The Neonatal Expiratory Valve is 

also working with Evita

®

 V-Series. 

For manual reprocessing of the

 

Neonatal Expiratory Valve please 
refer to Babylog guideline.

This guideline does not replace or amend the instructions for use. Please always refer to the 
reprocessing instruction in the latest instruction for use of the components. The instructions 
for use and the hospital guidelines and policies always have to be observed.