Disassembly
/Assembly
Flow Sensor
Open the flap by lifting the lower
edge upwards.
Close the flap.
Push the flow sensor as far as it
possible go to the left.
Push the flow sensor to
the right into the socket.
Remove the expiratory valve
from the fitting.
Turn the locking ring as far as
possible to the left.
Remove and empty the water
trap container.
Connect the water trap container.
Remove the diaphragm.
Fit the flow sensor
into the socket.
Remove the flow sensor
from the socket.
Remove the sensor sleeve
from the expiratory valve.
91
03
106
|
19.11-1
|
A
W
|
LE
|
Subject
to
modifications
|
©
2019
Dr
äg
er
w
er
k
A
G
&
C
o.
K
G
aA
Attach the diaphragm onto the
edge of the expiratory valve housing.
Make sure that it fitts properly.
Attach the flow sensor sleeve
on the expiratory valve.
Turn the locking ring as far
as possible to the right
until it clicks into place.
Turn the locking ring as far as
possible to the left, push the
expiratory valve into the fitting.
Dräger Evita
®
V-Series
Flow Sensor and Expiratory Valve
Disassembly/Assembly
Disassembly
/Assembly
Expiratory Valve*
* The Neonatal Expiratory Valve is
also working with Evita
®
V-Series.
For manual reprocessing of the
Neonatal Expiratory Valve please
refer to Babylog guideline.
This guideline does not replace or amend the instructions for use. Please always refer to the
reprocessing instruction in the latest instruction for use of the components. The instructions
for use and the hospital guidelines and policies always have to be observed.