![Dräger D-Vapor Instructions For Use Manual Download Page 85](http://html.mh-extra.com/html/drager/d-vapor/d-vapor_instructions-for-use-manual_2526743085.webp)
Instructions for use D-Vapor/D-Vapor 3000
85
Problem solving
Plug-in adapter problems
Fault Cause
Remedy
Locking lever does not engage
in control dial when disconnect-
ed.
Control dial still set to
0
.
Engage control dial in the
T
po-
sition.
Locking lever cannot be swung
out of the control dial.
Control dial set to
0
or
≥
2
Vol%.
During a preceding transport,
control dial may have been set
to
0
or
≥
2
Vol%.
Engage control dial in the
T
po-
sition.
Vaporizer cannot be disconnect-
ed.
Control dial not set to
T
.
Engage control dial in the
T
po-
sition.
Interlock still engaged.
Disengage Interlock.
Locking lever cannot be swung
back into control dial. Locking
device between plug-in adapter
and plug-in connector is
jammed.
Remove locking cap on top of
locking lever shaft.
Loosen screw in the shaft with
3 mm Allen key screwdriver.
The vaporizer can now be dis-
connected.
Repair by specialized service
personnel.
Plug-in adapter not fitting flush
on plug-in connector seals.
Leak in unit.
Locking lever not engaged in
control dial, as it is set at
0
or
≥
2
Vol%.
Set control dial to
T
and engage;
insert pin on locking lever into
slot on control dial and engage.
Engagement mechanism on
plug-in adapter or plug-in con-
nector damaged.
Excessive force used may lead
to jamming or problems when
disconnecting. Repair by spe-
cialized service personnel.
Locking lever was turned to the
left before connection.
Remove vaporizer (control dial
in position
T
); engage locking le-
ver in control dial; reconnect va-
porizer.
O-rings on plug-in adapter miss-
ing.
Fit O-rings.
Extra O-ring on a pin on the
plug-in connector or foreign
body between plug-in connector
and plug-in adapter.
Remove O-ring or foreign body.
Cable trapped between vaporiz-
er and anesthesia machine.
Correct the cable routing; ob-
serve the instructions for use for
the anesthesia machine.
Summary of Contents for D-Vapor
Page 14: ...14 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 15: ...Instructions for use D Vapor D Vapor 3000 15 Use Use Intended use 16 Environment of use 16 ...
Page 26: ...26 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 76: ...76 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 88: ...88 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 114: ...114 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 116: ...116 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 119: ...Instructions for use D Vapor D Vapor 3000 119 This page has been left blank intentionally ...