Problem solving
78
Instructions for use D-Vapor/D-Vapor 3000
Problems
Possible faults are listed in the following tables. The
tables will help to identify the underlying cause and
to quickly remedy it using the measures provided.
The different causes and remedies should be
worked through in the order listed until the fault has
been resolved.
Alarm cases
Priority
Alarm indication
Cause
Remedy
Warning
All LEDs flashing.
Equipment fault
Take equipment out of ser-
vice. Set control dial to the
0
position. Disconnect the
mains power supply. Re-
move the vaporizer from
the anesthesia workstation
if possible.
Repair by specialized ser-
vice personnel.
Warning
Green
Operational
LED
flashes.
Red
No Output
LED flash-
es.
Dosing request during
warm-up phase
Malfunction
Set control dial to the
0
po-
sition. Wait until the green
Operational
LED is lit con-
tinuously.
Warning
Red
No Output
LED flash-
es.
Yellow
Battery
LED flash-
es.
End of battery operation
Set control dial to the
0
po-
sition. Change to a vaporiz-
er that does not require a
mains supply.
Summary of Contents for D-Vapor
Page 14: ...14 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 15: ...Instructions for use D Vapor D Vapor 3000 15 Use Use Intended use 16 Environment of use 16 ...
Page 26: ...26 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 76: ...76 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 88: ...88 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 114: ...114 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 116: ...116 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 119: ...Instructions for use D Vapor D Vapor 3000 119 This page has been left blank intentionally ...