Instructions for use D-Vapor/D-Vapor 3000
43
Preparation
When the filling level has reached the refill mark, up
to 240 mL of the desflurane may be refilled using an
anesthetic agent bottle. This is indicated by the
yellow
Fill Up
LED lighting. A short tone sounds.
Vaporizer states for filling
The vaporizer can be filled in the following states:
– when switched off (power cable not in the power
socket)
– during the warm-up phase
– during operation
The filling procedure takes a little longer when
the vaporizer is warmed up. The anesthetic
agent must have reached ambient temperature
before filling. With a warmed-up vaporizer, cool
anesthetic agent or tight anesthetic agent
bottles, the filling time is increased. Filling time,
see page 102.
Vaporizer is connected to the anesthesia
workstation
Filling during operation: The fresh-gas flow can
remain set.
Vaporizer not connected to the anesthesia
workstation
The control dial remains engaged in the
T
position
(B).
WARNING
Risk of patient injury
Filling during battery operation shortens the
battery operating time.
Never fill the vaporizer models D-Vapor and
D-Vapor 3000 during battery operation.
WARNING
Risk of patient injury
Watch the filling level in the sight glass.
Fill the vaporizer to the maximum mark only.
When the filling level is below the minimum
mark, the vaporizer must be refilled.
CAUTION
Health hazard
Anesthetic agent may spurt if the filling device is
contaminated by dirt particles.
Keep the filling device clean. Dirt particles must
not get into the filling device.
NOTE
Before initial use or after long periods of storage,
warm up the Vaporizer until the green
Operational
LED is permanently lit. Then fill the vaporizer to the
maximum mark.
NOTE
The filling level in the sight glass will only be shown
correctly if the vaporizer is standing or hanging
vertically.
NOTE
Vaporizers that are supplied by a pharmaceutical
company and identified accordingly are intended
for use only with the anesthetic agent from that
company. Only use the filling adapter provided for
that particular filling system.
060
0
B
Summary of Contents for D-Vapor
Page 14: ...14 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 15: ...Instructions for use D Vapor D Vapor 3000 15 Use Use Intended use 16 Environment of use 16 ...
Page 26: ...26 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 76: ...76 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 88: ...88 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 114: ...114 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 116: ...116 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 119: ...Instructions for use D Vapor D Vapor 3000 119 This page has been left blank intentionally ...