67
Service
– There is no visible damage to device or power cables
4
Observe the instructions for use and check that all components and accessories
needed to use the product are present.
5
Check the electrical safety in accordance with the IEC 62353 or IEC 60601
standard.
12.5
Maintenance
The following table shows the preventive maintenance intervals. This table is not
related to warranty periods and no warranty is implied by the intervals listed.
The BiliLux LED panel does not need to be exchanged during the expected service
life of the device.
CAUTION
Risk of injury
The use of LED lights that are not approved by the manufacturer can affect the
safety and effectiveness of the phototherapy.
►
Use only LED lights recommended by Dräger. Contact DrägerService for further
information.
12.6
Maintenance instructions
Following any maintenance procedure, perform the functional checkout procedure
on page 38.
12.6.1
Exchanging the clock battery
1
Remove the back cover from the phototherapy light.
2
Remove the old clock battery.
3
Install the new clock battery (CR2032).
4
Reinstall the back cover on the phototherapy light.
5
Set the date and time.
12.6.2
Calibrating the radiometer
Calibration must be performed by specialized service personnel. Package the
radiometer in the protective case it was shipped in, and send it to DrägerService.
Component
Interval
Measure
Target group
Clock battery
Every 6 years
Exchange
Service personnel
Radiometer
Every 2 years
Calibrate
Specialized service
personnel
Spring arm joints
Every 1 year, or as
required
1)
1) When the spring arm no longer maintains its position
Tighten
Users or service
personnel
Summary of Contents for BiliLux
Page 2: ...2 Instructions for use BiliLux This page has been left blank intentionally ...
Page 6: ...6 Instructions for use BiliLux This page has been left blank intentionally ...
Page 8: ...8 Instructions for use BiliLux This page has been left blank intentionally ...
Page 18: ...18 Instructions for use BiliLux This page has been left blank intentionally ...
Page 26: ...26 Instructions for use BiliLux This page has been left blank intentionally ...
Page 30: ...30 Instructions for use BiliLux This page has been left blank intentionally ...
Page 36: ...36 Instructions for use BiliLux This page has been left blank intentionally ...
Page 54: ...54 Instructions for use BiliLux This page has been left blank intentionally ...
Page 58: ...58 Instructions for use BiliLux This page has been left blank intentionally ...
Page 64: ...64 Instructions for use BiliLux This page has been left blank intentionally ...
Page 70: ...70 Instructions for use BiliLux This page has been left blank intentionally ...
Page 84: ...84 Instructions for use BiliLux This page has been left blank intentionally ...
Page 88: ...88 Instructions for use BiliLux This page has been left blank intentionally ...
Page 89: ...Instructions for use BiliLux 89 This page has been left blank intentionally ...