![Dräger Babylog VN Series Manual Download Page 9](http://html1.mh-extra.com/html/drager/babylog-vn-series/babylog-vn-series_manual_2526409009.webp)
Supplement
|
Evita V series / Babylog VN series
9
Service
|
English
7 Service
7.1 Maintenance
7.2 Battery
maintenance
7.2.1
Information on battery maintenance
To achieve the maximum life span of the batteries, the following actions are
required:
Maintenance:
●
Observe the maintenance intervals. Batteries are wearing parts. The
replacement intervals depend on the utilization.
During operation:
●
Observe the ambient conditions (see "Technical data", page 13).
●
After battery operation, connect the device to the mains power supply. Fully
charge the batteries. Observe the required charging time.
!! 251
Battery low
The remaining calculated
operating time of the battery
is less than 10 minutes.
Connect the device to the
mains power supply.
!
100
Battery test in progress
The battery test was started
by the user but it was not
completed.
Wait until the battery test is
completed, before using the
device for patient transport.
!
100
Consider battery test
The interval for the battery
test has been exceeded.
Perform the battery test.
!! 210
Internal battery activated
The power supply unit
PS500 is depleted. The
device is supplied with power
from the internal battery.
Charge the power supply unit
PS500 as soon as possible.
The power supply from the
internal battery is limited.
To downgrade the alarm,
touch the "Alarm reset" but-
ton and confirm with the
rotary knob.
Component
Interval
Measure
User group
Batteries
Every 3 months
Check the capacity with
the battery test (see
"Battery test", page 10).
Service personnel
Replace if needed
Specialized service per-
sonnel
Internal battery (NiMH)
PS500 power supply
unit (VRLA)
Every 2 years or
depending on the
results of the battery test
Replace
Specialized service per-
sonnel
Power supply unit
PS500 (LFP)
Every 4 years or
depending on the result
of the battery test
Priority
Alarm
Cause
Remedy
Summary of Contents for Babylog VN Series
Page 21: ...Evita V Babylog VN 21 3 3 1 3 1 1 25 40813 1 2 3 4 2 3 1 4...
Page 22: ...22 Evita V Babylog VN 4 4 1 4 1 1 39417 1 2 3 1 2 3...
Page 24: ...24 Evita V Babylog VN 4 1 2 26 24 4 1 3 32 36 27 4 1 4 PS500 32...
Page 25: ...Evita V Babylog VN 25 5 5 1 5 1 1 27 28 5 1 2 PS500 PS500...
Page 26: ...26 Evita V Babylog VN 6 6 1 210 PS500 PS500 160 200 201 127 100 254 5...
Page 27: ...Evita V Babylog VN 27 7 7 1 120 100 250 10 251 10 3 28 NiMH PS500 VRLA 2 PS500 LFP 4...
Page 28: ...28 Evita V Babylog VN 7 2 7 2 1 32 32 5 5 7 2 2 GS500...
Page 33: ...Evita V Babylog VN 33 45 85 6 8 2 ISO 80601 2 12 2020 22 C 71 6 F 50 GS500 GS500 8 1 continued...
Page 36: ...36 Evita V Babylog VN NiMH VRLA LFP 12 Ah 2 1000 Ah 2 5 A GS500 9 1 3 25 49 0 24...
Page 39: ...Evita V Babylog VN 39 10 5 GS500 GS500 GS500...
Page 178: ...178 Evita V Babylog VN 3 3 1 3 1 1 182 40813 1 2 3 4 2 3 1 4...
Page 179: ...Evita V Babylog VN 179 4 4 1 4 1 1 39417 1 2 3 99 100 75 98 50 74 1 2 3...
Page 180: ...180 Evita V Babylog VN PS500 PS500 192 5 10 25 49 0 24 40757 1 2 PS500 3 4 3 1 2 4 3 4...
Page 181: ...Evita V Babylog VN 181 4 1 2 183 181 4 1 3 Te 189 193 184 4 1 4 PS500 Te 189...
Page 182: ...182 Evita V Babylog VN yp 5 yp 5 1 5 1 1 184 185 5 1 2 PS500 PS500...
Page 183: ...Evita V Babylog VN 183 6 6 1 C 160 200 201 210 PS500 PS500 254 5 100 127 250 10 251 10...
Page 184: ...184 Evita V Babylog VN 7 7 1 100 120 3 185 NiMH PS500 VRLA 2 PS500 LFP 4 C...
Page 185: ...Evita V Babylog VN 185 7 2 7 2 1 Te 189 Te 189 5 5 7 2 2...
Page 193: ...Evita V Babylog VN 193 NiMH VRLA LFP 12 2 1000 2 5 A GS500 9 1 3 25 49 0 24...
Page 195: ...Evita V Babylog VN 195 9 1 4 183 PS500 PS500 PS500 PS500 GS500 41052 36 36 6 12 18 24 1 2 2 1...
Page 196: ...196 Evita V Babylog VN 10 5 GS500 GS500 GS500...