150
Instructions for use Babyleo TN500 SW 1.0n
Transfer
9 Transfer
9.1
Intrahospital patient transfer with switched off warming
therapy
This chapter describes how to move the patient in the device into another room.
9.1.1
Overview of intrahospital patient transfer with switched off
warming therapy
The patient can be moved within the hospital in the device, e.g., from the labor and
delivery room to the neonatal intensive care unit. warming therapy is interrupted
during this time. The patient is not supplied with heat and humidified air.
The device must be set to a low height prior to moving the patient so that the red
hatched label on the lifting column is no longer visible. At this height, the device can
be moved without the radiant warmer hitting a door frame, for example.
Fig. 79
Label on the lifting column
9.1.2 Transferring
the
patient
This chapter describes how to prepare the device for transfer and how to transfer
the patient together with the device.
Prerequisites:
–
The device is in operation.
–
The device height adjustment mechanism is not locked.
WARNING
Risk of patient injury due to interruption of warming therapy
When the patient is moved to another room in the device, the device is switched off
during the transfer.The warming therapy is interrupted and the skin temperature is
not measured.
►
During the transfer, monitor the patient's body temperature with a separate
measuring device.
►
Make sure that the patient is sufficiently warmed during the transfer.
28226
Summary of Contents for Babyleo TN500
Page 2: ...2 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Page 212: ...212 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Page 214: ...214 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Page 221: ...Instructions for use Babyleo TN500 SW 1 0n 221 This page has been left blank intentionally...