Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
49
Assembly and preparation
Pressure reducers
Pressure reducers have an internal pressure release valve. If a fault occurs, gas
may escape into the ambient air. Personal injury or property damage may result as
a consequence.
►
Do not block or cover the pressure release valve.
4.1.3 Mechanical
safety
Accessories
If the weight of the accessories is unevenly distributed on the device or exceeds the
permissible limits, the device may tip over. The end stops of the support arms in use
must be functional, otherwise the arms may swing unchecked.
►
Distribute the weight evenly.
►
Pay attention to the maximum weight on each support arm.
►
Check whether the optional counterweight in the lower section of the trolley is
required (see "Compact version with counterweight", page 54).
►
Check the functional integrity of the support arm end stops after the following
activities:
– After fitting accessories
– After transporting the device
Trapping of body parts
Movable components or attached parts may cause crushing due to trapping. Pay
special attention to edges, movable parts, and corners when working with the
following components:
►
Breathing system cover
►
Drawers
►
Folding table extension
►
Pull-out writing tray
►
Support arms for mounted devices
►
Accessories such as gas cylinders, vaporizers, CLIC absorbers, and CLIC
adapters
Accidental movement of the trolley
An unbraked device may accidentally move during operation. As a result, user and
patient could be put at risk.
►
Prevent this by actuating the central brake or the castor brakes, and check their
function.
Strangulation
Negligent placement of hoses, cables, and similar device components can put the
patient at risk.
►
Use particular caution when establishing connections to the patient.