Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
31
Overview
3.2 Functional
scope
3.2.1
Product variants, options, and accessories
Some device functions are available as an option and consequently are only
available on appropriately equipped devices. Not all product variants or options are
available worldwide.
The device is intended for use with the options and accessories listed in the
associated list of accessories.
Hardware options
Hardware options are listed in the following table:
Name
Description
Compact version (model
designation A300, A350)
– Version with small trolley for environments of use
with constricted space
– 1 large drawer
– Central brake
– Available with plug-in connectors for 1 or 2 vaporiz-
ers
– The work surface can be enlarged with a folding
table extension on the side (option).
– When heavy workplace components, monitors, or
syringe pumps are attached, a counterweight is
required for increased tipping stability (option).
Large version (model des-
ignation A300 XL,
A350 XL)
– Version with large trolley for normal OR environ-
ments with adequate space
– 1 large and 2 small drawers
– Available with plug-in connectors for 2 or 3 vaporiz-
ers
– Can be fitted with a pull-out writing tray and a fold-
ing table extension on the side (option).
Mechanically controlled
gas mixer (model desig-
nation A300)
– The fresh-gas delivery is adjusted by means of
manually operated needle valves.
– The individual fresh-gas flows are measured elec-
tronically and displayed on both the status display
and the screen.
– The total fresh-gas flow is indicated on the inte-
grated total flow tube.
– Available as a 2-gas version (O
2
/Air) or a 3-gas
version (O
2
/Air/N
2
O)
– Available with O
2
flowmeter for O
2
insufflation
(option)