Reprocessing
13.11.2.5
Fitting the flow sensors and the ports
1. Insert the expiratory flow sensor (1).
2. Push in the expiratory port (2). Tighten the knurled nut.
3. Insert the inspiratory flow sensor (3).
4. Push in the inspiratory port (4). Tighten the knurled nut.
13.11.2.6
Fitting the CLIC adapter (optional)
1. Press the release button (1). The holder (2) flips open.
2. Remove the cleaning plate (3) from the adapter.
3. Close the holder (2) so that it engages.
4. Screw the CLIC adapter (4) on to the lower part of the breathing system.