Reprocessing Fabius family
27
English
After reprocessing
Assembling and fitting device-specific components
The device-specific components must be attached
to the device and assembled if necessary.
Install and fit the following components:
– Ventilator parts
– APL valve
– Inspiratory valve and expiratory valve
– Compact breathing system
– Flow sensor
– Exhaust port
– Breathing bag arm
– CLIC adapter
– CO
2
absorber
– O
2
sensor
– Hoses and cables
– Anesthetic gas receiving system
Prerequisites:
– All components have been reprocessed and
dried.
Installing the ventilator parts
1
Open the ventilator door (A) with the attached
ventilator unit.
2
Insert the piston diaphragm (B) into the
ventilator. Make sure that the Dräger inscription
on the bottom of the piston diaphragm is legible
from above.
3
Fit the ventilator lid (C). Close the 3 clamps (D).
4
Connect the pressure sensor line (E) of the
ventilator chamber to the corresponding
connector on the ventilator lid.
5
Close the ventilator door with the attached
ventilator unit.
Fitting the APL valve
1
Fit the APL valve vertically on the compact
breathing system.
The mark (A) must point in the direction of the
user when the compact breathing system is
installed.
2
Tighten the knurled nut (B) clockwise.
062
A
E
B
D
C
006
A
B