Reprocessing Fabius family
17
English
Surface disinfection with cleaning
Components:
– Device surface including screen
– Trolley and holders
– Removable parts of the main device
Prerequisites:
– The surface disinfectant has been prepared in
accordance with the manufacturer's
instructions.
– The manufacturer's instructions, e.g., regarding
shelf life or application conditions, are
observed.
– An uncontaminated, lint-free cloth soaked in
surface disinfectant is used for the cleaning
surface disinfection.
Cleaning
1
Wipe off obvious soiling with a disposable cloth
soaked in surface disinfectant. Dispose of the
cloth.
2
Wipe all surfaces. After that, there must no
longer be any soiling visible.
Surface disinfection
3
Wipe cleaned surfaces again to visibly wet all
surfaces to be disinfected with surface
disinfectant.
4
Wait for the surface disinfectant contact time.
5
At the end of the contact time, moisten a new,
uncontaminated and lint-free cloth with water
(at least drinking water quality).
6
Wipe all surfaces until no remains of the surface
disinfectant, such as foam residues or streaks,
are visible.
7
Wait until the surfaces are dry.
8
Check the surfaces for visible damage and, if
necessary, replace the product.
Components
Surface disinfectant Manufacturer
Concentration
Contact time
– Device surface includ-
ing screen
– Trolley and holders
– Removable parts of
the main device
Dismozon plus
Bode
Chemie
1.6 %
15 min
WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
– Damage to the device
– Electric shock
– Device malfunctions
Ensure that no liquid penetrates the device.