Dräger Medical AG & Co. KGaA
Service Mode Savina Version 3.0
Page 86
6HUYLFH0RGH
,QWHUQDO6HUYLFH0RGH
7HVW6WHS
Test step 10 is used to display the expiratory pressure "Pawexp" (S4.1) and the inspiratory
pressure "Pawinsp" (S4.2). The maximum permissible pressure is 105 mbar. The external
pressure measuring device indicates the opening pressure of the safety valve D1 (Target
value = 102 mbar to 110 mbar).
The blower delivers a flow 30 L/min; the expiratory valve is "sealed" with 120 mbar.
Connect (short-circuit) the inspiratory socket and the expiratory socket with a ventilation
hose and connect an external pressure measuring device to perform this test step.