Dräger Medizintechnik GmbH
Function Description Savina Version 2.0
Page 357
)XQFWLRQ'HVFULSWLRQ
3ULQFLSDOFRPSRQHQWV
'LVSOD\
The display shows the patient parameters and the warning messages. The display is actuated by
the Control PCB; the display backlighting is actuated by the Front Panel PCB. The display
resolution is 240 x 128 pixels. The voltage supply is +5 V and —15 V.
0HPEUDQHNH\SDG
The membrane keypad contains the keys, with associated LEDs, to operate Savina.
&RQWURONQRE
The control knob is used to set and confirm changeable ventilation parameters. The control knob
transmits square signals to the Front Panel PCB as it rotates, and the signals are then evaluated
by the Control PCB. The voltage supply is +5 V.