![Dräger Medical Narkomed GS Setup And Installation Manual Download Page 24](http://html1.mh-extra.com/html/drager-medical-0/narkomed-gs/narkomed-gs_setup-and-installation-manual_2527076024.webp)
3
Setup and Installation
3-14
Narkomed GS Setup and Installation Manual
Draeger Medical, Inc.
Installing the
Ultrasonic
Flow Sensor
1. Slide the flow sensor down onto the bracket attached to the expiratory
valve mount.
2. Install the connector hose on the threaded port of the flow sensor, and
join the other end of the hose to the expiratory valve. Ensure that the
expiratory valve is oriented as shown in the illustration below.
3. Connect the sensor plug to the VOLUME SENSOR interface on the
rear panel of the ventilator box.
SU00009A
FLOW SENSOR
BRACKET
ABSORBER
ASSEMBLY
EXPIRATORY
HOSE TERMINAL
EXPIRATORY VALVE
CONNECTOR
HOSE
ULTRASONIC
FLOW SENSOR
SENSOR
PLUG
2
Summary of Contents for Narkomed GS
Page 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 10: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 40: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 46: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 50: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 52: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 90: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 91: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...