Wrist Digital Blood Pressure Monitor 

Instruction Manual 

 

 

 

 

 

 

 

 

 

 

 

 

 

Wrist cuff is suitable for wrist circumference of 13.5to 21 cm

 

 

Summary of Contents for Wrist Digital Blood Pressure Monitor

Page 1: ...1 Wrist Digital Blood Pressure Monitor Instruction Manual Wrist cuff is suitable for wrist circumference of 13 5to 21 cm ...

Page 2: ...e set date 4 3 Language selection Setting the time and date and user selection 5 Carrying out a measurement 5 1 Before the measurement 5 2 Common sources of error 5 3 Fitting the cuff 5 4 Measuring procedure 5 5 Discontinuing a measurement 5 6 Memory storage and recall of the measurements 5 7 Novel features Positioning Detection movement detection in the measuring 5 8 Usage Instruction in Position...

Page 3: ...lication of the cuff and its pressurization on any limb where intravascular access or therapy or an arteriovenous A V shunt is present because of temporary interference to blood flow and could result in injury to the patient Do not let the cuff and its pressurization on the arm on the side of a mastectomy Pressurization of the cuff can temporarily cause loss of function of simultaneously used moni...

Page 4: ...ve 90 mmHg and or the systolic blood pressure is over 160 mmHg In this case please consult your doctor immediately Long term values at this level endanger your health due to the associated advancing damage to the blood vessels in your body Should the systolic blood pressure values lie between 140 mmHg and 160 mmHg and or the diastolic blood pressure values lie between 90 mmHg and 100 mmHg likewise...

Page 5: ...d pressure too high between 140 and 160 Between90and 100 Seek medical advice Blood pressure far too high between 160 and 180 Between100and 110 Seek medical advice Blood pressure dangerously high Higher than 180 Higher than 110 Urgently seek medical advice Further information If your values are mostly standard under resting conditions but exceptionally high under conditions of physical or psycholog...

Page 6: ...6 3 The various components of the blood pressure monitor ...

Page 7: ...lection Bilingual of English and Hindi There are two languages English and Hindi Press the time button for around 3 seconds uses icon will shining Then you just need to press time button several times until entering the language selection mode during this process you will see the year icon month icon time icon and switchable of mmHg and kPa icon shining just ignore them and press time button Defau...

Page 8: ... please proceed as follows 1 Press the TIME button for at least 3 seconds firstly user icon blink Then press TIME button again the display now indicates the set year during which the four characters blink 2 The correct year can be entered by pressing the MEMORY button 3 Press the TIME button again The display now switches to the current date during which the first character month blinks 4 The corr...

Page 9: ...he TIME button selects pressure unit 12 Now after all settings have been made press the TIME button once again The date is briefly displayed and then the time The input is now confirmed and the clock begins to run Further Information With each press of the button TIME MEMORY one input is made e g switching over from hours to minutes mode or altering the value by 1 However if you keep the respectiv...

Page 10: ...ll be measured Each 15cm difference in height results in a measurement error of 10mmHg A loose cuff causes false measurement values With repeated measurements blood accumulates in the respective wrist which can lead to false results Correctly executed blood pressure measurements should therefore first be repeated after a 5 minute pause or after the arm has been held up in order to allow the accumu...

Page 11: ...stole 118 Diastole 73 Pulse 75 The measurement results are displayed until you switch the device off If no button is pressed for 3 minutes the device switches automatically off to save the batteries 5 5 Discontinuing a measurement If it is necessary to interrupt a blood pressure measurement for any reason e g the patient feels unwell the ON OFF button can be pressed at any time The device then imm...

Page 12: ...ection On1 On2 and OFF On1 mode Positioning indicator detects whether or not the device is in the proper position in ON1 mode even if the device is not positioned properly and the Positioning indicator is orange the device is still workable an Orange Positioning indicator provides just a reminding of an improper wrist position or movement detected during measuring On2 mode the device is not positi...

Page 13: ...ion error ERR 5 The measured readings indicated an unacceptable difference between systolic and diastolic pressures Take another reading following directions carefully Contact you doctor if you continue to get unusual readings ERR8 Pressure in cuff is over 290mmHg Other possible malfunctions and their elimination If problems occur when using the device the following points should be checked and if...

Page 14: ... subjected to strict clinical tests by which the computer program used to measure the blood pressure values was tested by experienced specialist doctors in Germany The same computer program is used in every individual device and has thus also been clinically tested The manufacture of the devices takes place according to the terms of the European standard for blood pressure measuring devices You mu...

Page 15: ...ions section Protect it from water and moisture extreme temperatures impact and dropping contamination and dust direct sunlight heat and cold The cuffs are sensitive and must be handled with care Only pump up the cuff once fitted Do not use the instrument close to strong electromagnetic fields such as mobile telephones or radio installations Do not use the instrument if you think it is damaged or ...

Page 16: ...2009 corrigendum 2010 Electrical compatibility Device fulfils the stipulations of the IEC EN 60601 1 EN 60601 1 2 The stipulations of the EU Guidelines 93 42 EEC for Medical Products Class IIa have been fulfilled 12 Technical specifications Measurement Procedure Oscillometric corresponding to Korotkoff method Phase I systolic Phase V diastolic Display Digital display Measuring range Pressure 30 to...

Page 17: ...ironment specified below The customer or the user of the Wrist Digital BP Monitor should assure that it is used in such an environment Electromagnetic Emissions IEC60601 1 2 Emission Test Compliance Electromagnetic Environment RF emission CISPR 11 Group 1 The Wrist Digital BP Monitor uses RF energy only for internal functions Therefore this RF emission is extremely weak and there is little chance ...

Page 18: ... dip in UT for 5 sec Not applicable Mains power quality should be that of a typical commercial or hospital environment If the user of the upper arm stlye requires continued operation during power mains interruptions it is recommended that the Wrist Digital BP Monitor be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m Not applica...

Page 19: ...should be considered If the measuredfield strength in the location in which the Wrist Digital BP Monitor is used exceeds the applicable RF compliance level above the Wrist Digital BP Monitor should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the Wrist Digital BP Monitor b Over the frequency range ...

Page 20: ... using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer Note1 At 80MHz and 800MHz the separation distance for the higher frequency range applies Note2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from ...

Page 21: ... forbid to abandon and disposal at will Manufacturer s name and address Inapplicable baby Attention consult accompanying documents ingress of water IP22 TUV NO Reading Instruction Book before use Type BF equipment For any query contact dr drtrustusa com ...