Dr. Mach LED 115 Instructions For Use Manual Download Page 1

Mach LED 115

 

Dr. Mach

 

Lamps and Engineering

 

 

59060001 

Edition 01 

02.08.2012 / Bak  Page 1/15 

 

 

Instructions for use 

 

MACH LED 115 

 

Spot Light with LED Technology 

 
 
 

 

 
 
 
 
 

LED 115 stand model

 ............................................................................. 

Item no. 1153101200

 

LED 115 wall model

 ................................................................................ 

Item no. 1153102100

 

 

Special fixations: 

LED 115 with fixation with table clamp

 ................................................. 

Item no. 1153102500

 

LED 115 with fixation clamp for supply rails

 ........................................ 

Item no. 1153102601

 

LED 115 with fixation for round and square tubes

............................... 

Item no. 1153102602

 

LED 115 with fixation plate

 .................................................................... 

Item no. 1153102604

 

 
 
 

 

Dr. Mach

 

GmbH u. Co., Flossmannstrasse 28, D-85560 Ebersberg 

Tel.: +49 (0)8092 2093 0, Fax +49 (0)8092 2093 50

 

Internet: 

www.dr-mach.com

, E-mail: 

[email protected]

 

Summary of Contents for LED 115

Page 1: ... no 1153102100 Special fixations LED 115 with fixation with table clamp Item no 1153102500 LED 115 with fixation clamp for supply rails Item no 1153102601 LED 115 with fixation for round and square tubes Item no 1153102602 LED 115 with fixation plate Item no 1153102604 Dr Mach GmbH u Co Flossmannstrasse 28 D 85560 Ebersberg Tel 49 0 8092 2093 0 Fax 49 0 8092 2093 50 Internet www dr mach com E mail...

Page 2: ...ctions for use page 6 3 1 Stand foot page 6 3 2 Stand tube page 6 3 3 Operating the light page 7 3 3 1 ON OFF switch page 7 3 3 2 Positioning page 7 3 3 3 Electronic light intensity control page 8 4 Cleaning page 8 4 1 Mobile stand page 8 4 2 Light head page 8 4 3 Protection disk page 8 5 Maintenance page 9 5 1 Periodical maintenance work page 9 5 2 Settings at the light arm page 9 5 3 Settings at...

Page 3: ...stand if repairs and alterations have been carried out by themselves or a company who can guarantee that the safety regulations have been observed The manufacturer is not liable for personal or material damages if the lamp or the stand is misappropriate or incorrectly operated or misused General instructions All Dr Mach stand lamps are supplied with all installation and connection parts and compon...

Page 4: ...rds with the lock washer 4 and cheese head screw M8 5 so that the ex tension arm 6 can still fit easily into the central base Insert the extension arm 6 in to the central base from above holding the cam catch plate 3 against from the oppo site side Insert the other 4 extension arms 7 10 in the sequence shown in the left figure into the central base The extension arms are held by the cams of the ca...

Page 5: ...ment Mark the position of the holes to be drilled in the wall using the wall bracket Drill the holes and insert dowels provided by the customer Dr Mach does not include dow els in the scope of supply For mounting the lamp see point 2 2 2 4 Table attachment Screw table clamp onto table For mounting the lamp see point 2 2 2 5 Rail attachment Attach light mounting in the desired position on the rail ...

Page 6: ...al S on the roller Unlock by lifting the pedal again 3 2 Stand tube The stand tube can be adjusted continuously in height and fixed in the precisely required posi tion needing only one hand The height is adjusted by lifting the slip button 15 The position of the light is automatically fixed on letting go As long as the slip button 15 is raised the ex tension tube can be pulled off or inserted into...

Page 7: ...OFF switch Push the switch 16 on the power supply housing to switch the light on and off 3 3 2 Positioning Use the clamping lever 17 on the power supply housing to position the light arm Use the handle 18 on the light head to position the light The flexible joint G holds the light head in the requested position 16 17 G 18 ...

Page 8: ...simply wiping with a damp cloth Conven tional cleaning agents can be used In case of disinfection only use disinfectants with less than 20 alcohol 4 2 Light head The lamp has a high quality surface which can be cleaned with conventional cleaning agents Only use disinfectant with less than 20 alco hol 4 3 Protection disk The protection disk 20 is made of a high quality plastic For cleaning use luke...

Page 9: ...itching on 5 1 Periodical maintenance work The following maintenance work tests has have to be done every six months check on defects in paint work check on fissures at plastic parts check on deformation of the suspension The following maintenance work tests has have to be done once a year check the function electrical safety test check the suspension 5 2 Settings at the light arm Adjusting the cl...

Page 10: ...ed arm mounted to stand Light head Mach LED 115 Mach LED 115 Central light intensity at a distance of 0 5 meters 60 000 Lux Colour rendering index Ra 95 Light field size 14 cm Colour temperature Kelvin 4500 K Electronic light intensity control at the light head 50 100 Number of LED s 7 Life span of the LED s 40 000 h Diameter of the light head 22 cm Remark The technical data are subject to fluctua...

Page 11: ...ironmental conditions Operation Min Max Temperature 10 C 40 C Relative atmospheric humidity 30 75 Air pressure 700 hPa 1060 hPa Transport storage Min Max Temperature 10 C 50 C Relative atmospheric humidity 20 90 Air pressure 700 hPa 1060 hPa References on the package Temperature range for transport and storage Athmospheric humidity for transport and storage Air pressure for transport and storage ...

Page 12: ...EN 60601 2 41 Dr Mach GmbH is certified according to DIN EN ISO 13485 2003 8 Disposal The OT lamp doesn t contain any dangerous goods The components of the OT lamp should be properly disposed at the end of its shelf life Make sure that the materials are carefully separated The electrical conducting boards should be submitted to an appropriate recycling pro ceeding The rest of the components should...

Page 13: ...est level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 2 kV for po...

Page 14: ...o the transmitter manufactur er and d is the recommended separation distance in metres m b Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies ...

Page 15: ... as recommended below according to the maximum output power of the communications equipment Separation distance according to frequency of transmitter m Rated maximum output of transmitter W 150 kHz to 80 MHz P d 17 1 80 MHz to 800 MHz P d 17 1 800 MHz to 2 5 GHz P d 34 2 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 17 1 17 2 33 10 3 69 3 69 7 38 100 11 67 11 67 23 33 For transmitters rated at a maxi...