21
7.2 Status lights
The three status lights (1–3) provide information on Dosell’s status and assist with
diagnosing technical faults.
Status light 1 – Power supply
Green
Dosell is turned on and connected to mains voltage.
Yellow
Dosell is turned on, but has no mains voltage and has
switched to battery operation. Dosell has limited
battery life – return to mains voltage as soon as
possible.
Red
Dosell is turned on, but has no mains voltage; it has
switched to battery operation and the battery level is
critically low. Shut down Dosell and replace batteries
urgently (see section
6.7.1 Replacing the batteries
).
Return to mains voltage as soon as possible.
Off (no light)
Dosell has shut down.
Status light 2 – Connection to cloud
Yellow
Dosell is trying to establish a connection to the cloud.
Green
Dosell has successfully established a connection to
the cloud and has sent information. The green light
goes out after a minute or so.
Red
Dosell has failed to establish a connection to the
cloud, probably because there is no mobile coverage.
Try changing Dosell’s location. If the problem
remains, contact your place of purchase.
Off (no light)
By default, this light is always off except when it is
trying to connect to or has just made contact with the
cloud.
Status light 3 – Dosell’s current status
Green
Ready to dispense the next sachet.