Dolomite 3200016 User Instructions Download Page 1

 

 
 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MITOS P-PUMP 
REMOTE CHAMBER 400 

 

USER INSTRUCTIONS 

Summary of Contents for 3200016

Page 1: ...MITOS P PUMP REMOTE CHAMBER 400 USER INSTRUCTIONS...

Page 2: ...HPT 487168127 356_v 1 0 Page 2 of 12 Contents 1 Mitos 400 Safety Information 5 2 System Set up 7 2 1 Introduction 7 2 2 Using the Mitos P Pump Remote Chamber 8 3 Mitos P Pump Control 11 3 1 Control 11...

Page 3: ...e fluidic connection to be made between the Mitos P Pump and a microfluidic system Flow Resistors F1 F3 F10 F30 F100 Part No 3200028 3200032 The Flow Resistor can be used to insert additional flow res...

Page 4: ...No 3200045 This Vessel Holder is designed to accommodate 3 micro tubes in the Mitos P Pump Included in the kit are compatible micro tubes with 1 5ml volume When used in conjunction with the Mitos P P...

Page 5: ...owable Temperature Range 0 50 C Do not subject the remote chamber to internal or external temperatures over 50 C or below 0 C The remote chamber may be used on a hotplate providing the plate surface t...

Page 6: ...tos 400 Safety Information The Mitos 400 remote chamber is manufactured and tested in accordance with Article 4 paragraph 3 of the pressure equipment directive 2014 68 EU and is classified under Sound...

Page 7: ...pressure gauge pressure on the Mitos P Pump which determines the flow rate of liquid from the Remote Chamber When the user vents or when the Mitos P Pump is switched off the pressure inside both chamb...

Page 8: ...e chamber lid with dip tube fitting Pneumatic connector 250ml Duran bottle Mitos P Pump not included Compressed gas or air supply not included Step 1 Unscrew the 4 thumb screws in the remote chamber l...

Page 9: ...nd tighten the 4 thumb screws evenly Do not completely tighten one thumb screw before tightening the others as this will result in uneven clamping force Ensure that all thumb screws are fully tightene...

Page 10: ...onnect the Remote Chamber to the Mitos P Pump Step 8 Set pressure on Mitos P Pump see Mitos P Pump user instructions to start pumping To stop pumping vent the chamber by pressing the back button on th...

Page 11: ...Control When operating with the Mitos P Pump Remote Chamber 400 it will take significantly more time for the chamber to reach the desired pressure due to the increased volume However the Mitos P Pump...

Page 12: ...SHPT 487168127 356_v 1 0 Page 12 of 12 Dolomite Microfluidics 3 Anglian Business Park Royston SG8 5TW UK T 44 0 1763 242491 E info dolomite microfluidics com W dolomite microfluidics com...