PATIENT QUICK START GUIDE
CALF CUFF APPLICATION
Wrap the cuff around the calf and secure the
hook and loop tabs to hold it in place. Make
sure the wrap is snug, but not too tight.
CUFF PLACEMENT
When both wraps are secured on your legs
they should look like the picture above.
TURNING THE DEVICE ON
When the wraps are secured on your legs
PRESS
and
HOLD
the
WHITE
power button
for about a second until the light is illuminated
on each unit.
USING THE DEVICE
Don’t be startled. The device will make a
“humming” sound when inflating and
squeezing your leg.
THIS
IS
NORMAL
. The
wraps will inflate once a minute. IF you
FEEL
air releasing around the legs, this is normal.
This is a function of the device to keep your
legs cool.
APPLICATION:
POWER INDICATORS:
What to do if the Pump
Alarms
Don’t worry, the alarms are
there to ensure the units are
operating correctly.
BATTERY CRITICAL
If battery voltage drops below a critical
level, cycling stops, audible alarm will sound,
and
UNIT
WILL
FLASH
YELLOW
. The Alarm will
continue for 30 seconds (unless the unit is
powered off) and automatically turn the unit
OFF
.
If the power button illuminates
GREEN
the units
are fully operational.
If the devices
DO
NOT
turn
ON
, or the
power button is “
YELLOW
” or “
FLASHING
YELLOW
”,
plug in the power cord immediately to the
electrical outlet.
IF both units are plugged in,
AND
turned
ON
the
power button will be
GREEN
and
the batteries WILL charge at the same time.
When the unit is turned
OFF
and plugged in the
battery is charging the
POWER BUTTON FLASHES RED.
Once the battery reaches full charge the
POWER
BUTTON WILL BE RED
and
REMAIN SOLID
For Application and Usage Questions Please Contact Customer Care:
877.475.7050
BEFORE USING THE DEVICE, PLEASE READ THE
INSTRUCTIONS FOR USE
COMPLETELY AND CAREFULLY. CORRECT APPLICATION IS VITAL TO THE PROPER FUNCTIONING OF THE DEVICE.
LOW PRESSURE OR LEAK
If the
ALARM
sounds and the
BLUE
LED
is
flashing, the pressure is too
LOW
. Make sure wrap
is attached snugly to the leg. Turn the Unit
OFF
and then back
ON
to reset the alarm. If the unit
continues to
ALARM
after this step,
DO
NOT
try to
fix the device.
ALARMS
:
USB PORT (PHYSICIAN USE ONLY)
LOW PRESSURE INDICATOR
POWER BUTTON
& BATTERY INDICATOR
CHARGING PORT
CHARGING YOUR VENAPRO
®
:
If you wear and operate the VenaPro
®
unit while charging
expect it to take approximately 4 hours to fully charge
(from a completely depleted battery.) If the unit is not in
use during charging, a complete charge (from a
completely depleted battery) will take approximately 3 hours.
PLEASE NOTE:
Units are not shipped fully charged. Fully
charge unit before use.
LEFT LEG SHOWN HERE
2
1
4
3