page 27 of 36
ANNA-LISA Manual - Version 1.1
CHAPTER 4: Operation
Amplitude
If the BNC input is active, the voltage of the MADI carrier is measured. The AES10
spec requires it to be between 300..600mV.
Code
Measured Voltage
LED Code
< 250 mV
LOW blinking red
OK OFF
HIGH OFF
250 .. 350 mV
LOW ON yellow
OK OFF
HIGH OFF
350 .. 550 mV
LOW OFF
OK ON green
HIGH OFF
550 .. 650 mV
LOW OFF
OK OFF
HIGH ON yellow
> 650 mV
LOW OFF
OK OFF
HIGH blinking red
For the SFP input, the RX power measurement of the SFP module is read out and
displayed in the same fashion as for the BNC input. If the SFP module has no DDM
capabilities, the ‘AMPLITUDE’ led blinks to indicate that this parameter cannot be
measured.
Parameter: AMPLITUDE
Values: LOW / OK / HIGH