2 • Description
Doc. 065207-02 9/08
11
2.1.2
Sample Cells and Rinse Cells
NOTE
contains part numbers for cells, bottles,
vials, and other accessories.
Sample Cells
Sample cells consist of a cell body and two interchangeable caps, which
are screwed onto each end of the cell body. The cell body and end caps
are made of the same material. Inside each end cap is a frit in the same
material as the cell, as well as a PEEK™ seal. During a run, the cell end
caps are compressed by the oven to form a tight seal between the caps and
the cell body.
Each cell end cap contains an external O-ring. Teflon® O-rings
(P/N 049457, pkg. of 50) are standard. Use Viton® O-rings (P/N 056325,
pkg. of 50) for dioxins and other high temperature applications.
Moves the cursor, in the direction of the arrow, to the next
selectable line on the display (if any).
When the cursor is in an editable field, pressing an arrow
button displays the next or previous parameter or numeric
value allowed for the field.
Note:
Pressing and holding down an arrow button moves the
cursor continuously through the allowed settings.
Cell Material
Sizes Available (mL)
Used For
Stainless steel
1, 5, 10, 22, 34, 66, and
100
Extractions with solvents
Zirconium
66 and 100
Extraction of basic or acidic
matrices, and extractions with
solvents
Always tighten the cell end caps by hand. Use of a wrench or other
tool can damage the cell, as well as the seals inside the cell end caps.
If Viton external O-rings are installed on the cell end caps, do not use
acetone or other ketones.
Button
Function
Table 2-1. ASE 150 Control Panel Button Functions (Continued)
Summary of Contents for ASE 150
Page 32: ...ASE 150 Operator s Manual 24 Doc 065207 02 9 08...
Page 64: ...ASE 150 Operator s Manual 56 Doc 065207 02 9 08...
Page 76: ...ASE 150 Operator s Manual 68 Doc 065207 02 9 08...
Page 114: ...ASE 150 Operator s Manual 106 Doc 065207 02 9 08...
Page 132: ...ASE 150 Operator s Manual 124 Doc 065207 02 9 08...