ASE 150 Operator’s Manual
134
Doc. 065207-02 9/08
C.2.4
Hydrocarbon Sensor Screen
Use the
HYDROCARBON SENSOR
screen to monitor the hydrocarbon
level at the ASE 150 installation site and to adjust the solvent vapor
threshold, if necessary.
Figure C-9. Hydrocarbon Sensor Screen
Parameter
Description
ACTUAL
Indicates the hydrocarbon level detected by the
hydrocarbon sensor.
THRESHOLD
Selects the solvent vapor threshold (10 to 10000). If
the threshold is exceeded, an error occurs. If
necessary, adjust the threshold to compensate for
high background levels at the installation site.
ACTUAL
THRESHOLD
0
100
HYDROCARBON SENSOR