C.2.3
Regulators Screen
Use the
REGULATORS
screen to monitor various pressure readings for the
ASE 150. You cannot edit any information on this screen.
NOTE Select the unit of measure for on-screen pressure
readings on the
SETUP
screen (see
NOTE The
COMPRESSION
and
AUTOSEAL
pressure
readings are zero unless a run is in progress.
Figure C-8. Regulators Screen
Parameter
Description
Specification
SYSTEM
Indicates the internal system pressure.
0.34
±
0.02 MPa
(50
±
3 psi)
COMPRESSION
Indicates the pressure applied when the oven is
compressed
0.90
±
0.03 MPa
(130
±
5 psi)
BOTTLE
Indicates the pressure applied to the solvent
reservoir
0.03 to 0.07 MPa
a
(4 to 10 psi)
a.
Nominally set for 0.04 MPa (6 psi). The pressure may increase during periods of system
inactivity, but will return to a value within the specification range as soon as a run begins.
AUTOSEAL
Indicates the pressure applied when the
AutoSeal air cylinder is actuated
0.14
±
0.01 MPa
(21
±
2 psi)
SYSTEM
COMPRESSION
BOTTLE
AUTOSEAL
50PSI
0PSI
6PSI
0PSI
Summary of Contents for ASE 150