B • Installation
Doc. 031842-01 5/02
B-7
4. Compare the pressure readings on the
REGULATORS
screen with the
values listed in the table below. If a pressure reading does not meet
specification, contact Dionex.
Field
Function
Pressure Spec
SYSTEM
Pressure applied to the system
0.34
±
0.02 MPa
(50
±
3 psi)
COMPRESSION
Pressure applied when the oven
is compressed
0.90
±
0.03 MPa
(130
±
5 psi)
BOTTLE
Pressure applied to the solvent
bottle
0.03 to 0.07 MPa*
(5 to 10 psi)
AUTOSEAL
Pressure applied when the
AutoSeal air cylinder is
actuated
0.10
±
0.01 MPa
(15
±
2 psi)
*This pressure is nominally set for 0.04 MPa (6 psi). The pressure may
increase during periods of system inactivity, but will return to a value
within the normal range as soon as an extraction begins.