DGT4X

www.

diniargeo

.com

QUICK START GUIDE V1

 

ENGLISH

Digital weight Transmitter with 4 channels

Summary of Contents for DGT4X

Page 1: ...DGT4X www diniargeo com QUICK START GUIDE V1 ENGLISH Digital weight Transmitter with 4 channels...

Page 2: ...C EXC 11 12 16 17 9 10 RS485 B A J1 13 14 15 GND RX TX RS232 JUMPER J1 QSG_ENG_DGT4X 1 Electrical scheme Internal use INPUT 12 48 Vdc OUTPUT 48Vac or 60Vdc 0 5 A max CONSUMPTION 4 5 W max For UL appro...

Page 3: ...eases digit Scroll down Increases digit Scroll up Enter the setup Selects digit to modify Enters a step Confirms Clears Exits a step no save Weighing mode Clears the displayed gross weight With short...

Page 4: ...on ch 7 Zeroing of the deal load pre tare zeroing ch 8 Load cell exclusion ch 14 Date and time setting with optional card only RS485 port configuration ch 13 Digital input configuration ch 11 Digital...

Page 5: ...29 SIG 28 SIG 25 SIG 24 EXC 23 REF 21 REF 20 SIG 19 SIG 18 26 EXC 22 EXC EXC 30 EXC 31 SIG 32 SIG 33 EXC 27 24 Vdc Earth 1 RL 1 6 RL 2 7 5 IN 1 4 COM 3 2 COM 8 GND IN 2 I 10 I 9 GND 15 A 16 14 TX 13 V...

Page 6: ...5 1 2 5 10 20 50 Set the maximum scale capacity 2 max 999999 Examples 7 Theoretical calibration Set the total load cells capacity up to 999999 Set the load cells sensitivity up to 999999 mV V cell1 mV...

Page 7: ...ro point acquisition 1 Unload the scale 2 Enter the zero step to adjust the zero point 5 Save adjustment Press key many times until save message will appear then press to confirm Sample weight acquisi...

Page 8: ...oard is locked When input is active transmitter reboots Emulation of key Emulation of key Emulation of key Emulation of key Emulation of key 12 Output setting See out 02 Setpoint on gross weight Setpo...

Page 9: ...n Modbus RTU ch 18 14 Broken laod cell exclusion for dependent channels systems No channel excluded Channel 1 excluded Channel 2 excluded Channel 3 excluded Channel 4 excluded If a load cell is broken...

Page 10: ...tion error Repeat the calibration checking that the capacity and division have been correctly set Er 39 Instrument not configured Transmitter needs to be configurated C er 36 Calibration error Check t...

Page 11: ...7 msb Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0 lsb Last command result Last command result Last command result Last command result Counting of processed commands Counting of processed commands Counti...

Page 12: ...ale 4 40221 Data Register DESCRIPTION Status register scale 1 40202 Bit 15 msb Bit 8 lsb No function Bit 7 msb Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0 lsb Tare PT 1 PT tare is active Tare 1 Tare is...

Page 13: ...04 Second parameter of the command Parameter is always expressed in absolute mode no decimals no sign 40005 EXAMPLE 1 For zeroing the weight on the scale 2 Set the command in byte 2 Byte Value 1 00 He...

Page 14: ...its products at any time without notice and without incurring any penalty We therefore recommend that you always check for any updates The individual in charge of the scale operation must ensure that...

Page 15: ......

Page 16: ...ted version on the www diniargeo com website or by contacting the Dini Argeo Customer Service Rev 30 01 2020 HEAD OFFICE Via Della Fisica 20 41042 Spezzano di Fiorano Modena Italy Tel 39 0536 843418 F...