17.2 AT START-UP
MESSAGE
DESCRIPTION
LED DISPLAY
LCD DISPLAY
Err. – 36
During the calibration some internal negative points have been
calculated:
-
the calibration point is less than the zero point.
-
the signal is negative (check the connections).
Err. – 37
During the calibration some internal points less than the minimum
value have been calculated:
-
the calibration point is equal to the zero point.
-
A capacity too high in relation to the division has been set.
Err. – 39
In alternation with
“no.CAL”
The converter points are less than the instrument’s internal
divisions; one must execute a TECHNICAL DEFAULT (
dFLt.t
parameter of the SET-UP environment) and carefully carry out
the calibration.
Err. – 39
Invalid calibration range number (there is the value 0 or values
greater than 3); one must execute a TECHNICAL DEFAULT
(
dFLt.t
parameter of the SET-UP environment), if not already
executed previously, and carefully carry out the calibration.
Err. – 40
The value 0 is in the “range 1” parameter of the calibration; one
must execute a TECHNICAL DEFAULT (
dFLt.t
parameter of the
SET-UP environment) and carefully carry out the calibration.
Err. – 41
The value 0 is in the “diV 1” parameter of the calibration; one
must execute a TECHNICAL DEFAULT (
dFLt.t
parameter of the
SET-UP environment) and carefully carry out the calibration.
Err. – 98
Upon start-up, the indicator is waiting for the data.
hW-Err
HARDWARE ERROR: software not compatible with the installed
hardware; the hardware expansion component is missing which
allows the software to function.
Err.C.0Y
ERROR SX
CELL Y
It is shown when there is an error in the communication with the X
scale – Y cell: check the connections and communication
parameters
(TECH.MAN.REF.)
.
ADC CONVERTITOR
ERROR
Possible indicator failure
Contact your dealer or authorized service center.
NOTE: if the Err – XX error appears after the indicator start-up, press the TARE key to enter in the SET-UP
environment.