3590EXTLWBCIO, CPWELW series indicator
E-LW_01_14.03_EN_U
45
15.4.1 SETPOINT ON THE GROSS WEIGHT
Functioning with hysteresis
(
rL.iSt
“HYSTERESIS” parameter set as
iSt.on
)
By selecting this functioning mode the relays function on the GROSS weight is activated; one enters two SETPOINTS for
each relay: a DISABLING one, which, when the gross weight is lower than it, disables the appropriate relay; an ENABLING
one, which, when the gross weight is greater or equal than it, it enables the appropriate relay. By pressing the
F3
key.
The SET POINT values are entered (DISABLING and ENABLING) for each configured relay:
-
The display shows "
S.1 on
" (relay 1 ENABLING SETPOINT): press
ENTER,
enter the weight value with the
keyboard and confirm with
ENTER
(use the
C
key to quickly zero the current value).
-
The display shows "
S.1 oFF
" (relay 1 DISABLING SETPOINT): press
ENTER,
enter the weight value with
the keyboard and confirm with
ENTER
(use the
C
key to quickly zero the current value).
-
Proceed in the same way for the following relays (if present).
-
When finished programming the setpoints, exit with the
C
key.
FUNCTIONING WITHOUT HYSTERESIS
(
rL.iSt
“HYSTERESIS” parameter set as
iSt.oFF
)
The operating mode is the same as the previous one, except that just one SETPOINT is entered for each relay (" S.1 on " );
for this purpose
the access to the " RELAY XX OFF " parameter is inhibited.
NOTES
-
If no relay has the “
GROSS
” functioning mode, the pressing of the
F3
key after the selection of the
<<SEt.Pnt>>
menu has no effect.
-
The DISABLING SETPOINT must be less or equal to the ENABLING ones; if a value greater than the ENABLING
one is entered and confirmed in the DISABLING SETPOINT, the indicator will set at 0 the setpoint, until a correct
value is entered.
If a value lower than the DISABLING one is entered and confirmed in the ENABLING SETPOINT, that value is
accepted, but the DISABLING SETPOINT is put at 0.
-
The 0 value is valid on both the setpoint values.
-
The SETPOINT check on the weight is active just during the weighing (in other words, not during the modification of
the SETPOINTS).
-
In case of weight not valid (local or from remote scale) all the relays will be disabled.
NOTE:
For the specification of all the relay functions, please see the
<< setup >>
→
<< outPut >>
→
<< r.ConF >>
→
<<rLFunC>>
step,
TECH.MAN.REF.
15.4.2 WEIGHT AT ZERO
By selecting this functioning mode the output function on the GROSS weight at 0 is enabled.
15.4.3 INSTABILITY
By selecting this functioning mode the output function on the unstable weight is enabled.