8.2 Conditions of malfunction on the device
Error
Cause
Solution
USB data saving error
1. Incorrect connection of
the USB card
Check correct insertion of the
USB card
2. Malfunctioning of the
USB card
Format the USB card and
retry.
Use another USB card
3. Incorrect
communication
between USB card
and device
Contact Customer Service
Prolonged historical
initialisation upon
power-up
1. First initialisation of the
historical analysis
Wait until completion of the
historical initialisation (a few
minutes) and run an analysis
(this must be successful).
X-
NIR ™ does not turn
on
1. Battery not connected
Check that the battery is
inserted inside the device;
Check that the battery is
inserted correctly.
2. No battery power
Check the charge level of the
battery;
Charge the battery.
3. HW problem
Contact Customer Service
Analysis errors
1. Incorrect analysis
procedure
Carefully follow the
instructions in the manual on
paragraph 5.7;
Clean the reading glass after
each analysis;
Check the correspondence
between the NIR family
selected and the forage that is
being analysed.
2. Incorrect loading of the
calibration file
Check the file loading
procedure on paragraph 5.12
of the manual.
Check correct functioning of
the USB card.
3. Analyzer malfunction
Check the status icon of the
device (list on paragraph 5.4
of the manual);
Perform a device Check
Hardware (instructions on
paragraph 5.9 of the manual);
Contact Customer Service