5.7 Analysis Execution with X-
NIR™ System
After turned ON the X-
NIR™ system and waiting for the following phases:
WARM-UP (at least 15 minutes) and CALIBRATION
please see icons in the
paragraph “States and Messaging of NIR analyzer internal system” for more details;
Setting of the customer information (CUSTOMER CODE
and CUSTOMER
INFORMATIONS
);
it’s possible to prepare the sample and starts to execute the analysis.
Select the family of sample to analyze from the list on the main page:
Please be sure to select the correct Component / NIR family before starting
the analysis on the forage sample because it represents the feed
components category which the sample to analyze belong.
The correct position for performing an analysis assumes that the device is perpendicular
to the area to be treated and is resting perfectly on the material to be analyzed, that is,
that the reading window is in contact with the element avoiding the formation of any
empty spaces which could alter the reading :