5.5 Initial optical control of the X-NIR
™ system
Every day, BEFORE starting to use the X-NIR
™
dinamica generale
analyzer, it
is advisable to perform the OPTICAL TEST via the CHECK CELL supplied as
an accessory.
CHECK CELL
particular sample in acetal resin for the X-NIR
™
tests.
The correct position to perform an OPTICAL TEST requires the device to be perfectly
supported on the check cell and perpendicular to the work surface:
Follow the steps below to perform an
OPTICAL TEST
.
1.
Position the CHECK CELL on the work surface with the white part pointing upwards, and
place the optical face of the X-NIR
™ device as shown in the figure.
Check cell
Work surface