![Dinamica Generale DG8000 Operator'S Manual Download Page 41](http://html1.mh-extra.com/html/dinamica-generale/dg8000/dg8000_operators-manual_2505533041.webp)
Configuration
INDICATOR DG8000 RevB0
39
E
N
GL
IS
H
Password 46 – Inverse Calibration with sample weight
This procedure allows to carry out the calibration of the weighing system on the basis of a
known weight that must be appropriately weighed on the medium in which the weighing
system is mounted.
Load the weight known inside the wagon before turn on the indicator!
1
Enter the value of the reference weight
placed in the wagon (in kg or pounds
depending Unit currently set size) with
↑
and
↓
buttons and confirm with the
ENTER key.
2
Confirm with the ENTER button to
proceed with the calibration.
3
Empty the wagon and press the
ENTER button to start the calibration
4
Wait for the scan job in progress.