Page 1: ......
Page 2: ......
Page 3: ...ith sub second resolution Rugged design for field or laboratory use Dimensions 6 2 x 3 7 x 1 3 Only requires 4 AA alkaline or Micro B USB Range of 1 to 775 Torr 29 92 to 0 59 inches Hg and mbar 2 Torr...
Page 4: ...or this user manual Remove protective boot Remove the battery cover and install 4 AA batteries or attach a Micro B USB Plumb Piezo vacuum sensor into the system to be measured taking care to differen...
Page 5: ...5...
Page 6: ...s would make it s accuracy at 10 Torr 2 10 or 20 of reading at an actual pressure of 10 Torr and at 200 Torr 2 200 or 1 of reading The most important thing about accuracy is knowing what you need for...
Page 7: ...vious there it may be time to replace the sensor The sensors are easily replaced and have excellent interchangeability Isolated sensors are immune to damage from corrosive processes and materials The...
Page 8: ...he top left figure is the maximum value in this range and the bottom left figure is the minimum value in this range The top right figure is the current or most recent value in the instrument Sometimes...
Page 9: ...ge in the pressure range of measurement which results in a quieter gauge Scientific The scientific format has two digits of precision with a base ten exponent following conventional usage For example...
Page 10: ...figuration menu and the PWR EXIT button turns the gauge on or off The battery indicator shown at full strength In practice this indicator changes based on the current level of pressure The numeric dis...
Page 11: ...mum value in this range and the bottom left figure is the minimum value in this range The top right figure is the current or most recent value in the instrument Sometimes this visualization is not use...
Page 12: ...module which interfaces with the Vacuum Gauge Application for remote real time vacuum monitoring via Apple Store at https appsto re us B1Ix5 i or Google Play Store at https play google com store apps...
Page 13: ...t to the sensor The sensor itself is cannot be calibrated but the sensor controller pair is See our blog on sensor interchangeability effects on accuracy for the Bullseye Precision Gauge Piezo for mor...
Page 14: ...actly 20 Torr then press the up and down buttons simultaneously so the unit reads exactly 20 Torr or 20000 0 Press the arrow button once more so the screen reads Press up down to Store then press up a...
Page 15: ...the select button until the screen reads Press up down to Store then press up and down button simultaneously Screen will flash black and calibration offsets will be saved Note If field calibration is...
Page 16: ...D Product Dimensions 6 2 x 3 7 x 1 3 Power 4 AA batteries or Micro B USB Compliance CE and RoHS Wetted Materials 316SS Numeric and graphical displays See your vacuum readings in either numbers or easy...
Page 17: ...y particular square inch This 14 696 PSI is equivalent to the pressure produced by a mercury column of approximately 29 92 inches high or 76 meters 3 4 of a yard or 760 millimeters of mercury Atmosphe...
Page 18: ...18...
Page 19: ...O Calibrated under actual vacuum against a NIST standard Patent Pending vacuum graphing displays vacuum and time data Visual graphing right on the display that identifies current system status Vacuum...
Page 20: ...SSED OR IMPLIED Limitation Of Liability You agree and acknowledge DIGIVAC shall have no liability to you whatsoever for any direct indirect punitive incidental special consequential damages arising ou...