DIATECH PHARMACOGENETICS EasyPGX qPCR instrument 96 User Manual Download Page 7

 
 

EasyPGX

® 

 qPCR instrument 96 

Cat.Nr: RT800-96 
 

7/21

 

 

                                                  

 2017/06 

 

 
 

SAFETY INFORMATION 
 

This product is designed for easy and reliable use in compliance with accepted safety standards. Use does not involve 
any risks if it is in accordance with the instructions contained in this document. However, misuse may result in damage to 
the  equipment  or  cause  health  hazard.  It  is  important  that  the  following  safety  precautions  are  read  and  understood 
before using the instrument. 
All  users  should  read  and  understand  this  user  manual  and  use  the  unit  only  in  accordance  with  the  instructions 
provided. Failure to follow the instructions may cause the protection provided by the instrument to be compromised. 
 

 
Pay particular attention to the paragraphs marked with this symbol. 

 

ELECTRIC SAFETY 

7.  Connect the instrument only to an appropriate voltage circuit (100-240 VAC). 
8.  Make  sure  that  the  power  cord  and socket  are easily  accessible  while  using  the  instrument  so that  the  instrument 

can be easily disconnected if necessary. 

9.  Do not connect the instrument to a socket without proper electrical grounding. 
10.  Do not touch any electrical switch or socket with wet hands. 
11.  Switch off the instrument before disconnecting it from the electrical socket. 
12.  Disconnect the instrument from the socket before any cleaning or maintenance work. 
13.  If liquid infiltrates the unit, disconnect the power socket and contact the supplier. 
14.  Do  not  connect  the  instrument  to  the  same  socket  where  other  instruments  with  high  electrical  consumption  (e.g. 

refrigerators or freezers) are connected. 

15.  Maintenance or repair activities that involve electrical components can only be performed by qualified personnel. 

FLUIDS AND REAGENT 

 

Fill  the  reaction  tubes/strips/plates  keeping  the  safety  distance  from  the  instrument  to  prevent  accidental  spills  of 
liquids on the instrument. 

 

Do not use the instrument to process inflammable, explosive or reactive substances. 

 

It  is  responsibility  of  the  user  to  apply  appropriate  safety  measures  if  potentially  pathogenic  material,  radioactive 
substances or other hazardous substances are used. 

 

Do not immerse the instrument in any liquid. 

 

DANGER OF BURNS 

 

Do not touch the thermoblock, the inner part of the lid or the reaction tubes/strips/plates while the instrument is 
operating. These areas can quickly reach temperatures greater than 50°C. Keep the lid closed until the temperature 
has fallen below 30°C. 

 

Do not use any material (tubes, foil, ...) that is not sufficiently stable at high temperatures (up to 120°C). 

OPERATIONAL ENVIRONMENT REQUIREMENTS 

 

The  fans  placed  on  the  hull  of  the  instrument  must  always  be  free  from  obstructions  and  free  to  operate.  Always 
leave at least 10 cm of space around the instrument. 

 

Keep the environment at a temperature between 20 and 30 °C and between 20 and 80% relative humidity. 

 

Do  not  use  the  instrument  in  environments  where  aggressive  or  explosive  volatile  mixtures  are  present.  Contact 
Diatech  Pharmacogenetics  support  if  you  need  to  use  the  instrument  in  non-conventional  environment 
(

[email protected]

). 

ELECTROSTATIC DISCHARGES 

Electrostatic discharges greater than 8000 volts can interfere with the functionality of USB ports on the instrument. It is 
necessary  to  handle  the  instrument  carefully  if  you  are  in  a  highly  electrostatic  environment  by  applying  appropriate 
antistatic measures before touching the instrument. ESD STM5.1- 1998 Class 3B. 
 
The 

“EasyPGX

®

 qPCR instrument 96

 complies with IEC 61326- 2-6 standards regarding electromagnetic compatibility 

for  IVD  medical  equipment.  The  instrument  has  been  designed  and  tested  for  CISPR  11  Class  A.  In  a  domestic 
environment  it  can  cause  radio  interference,  in  which  case  it  may  be  necessary  to  take  measures  to  mitigate  the 
interference. It is advisable to choose an appropriate electromagnetic environment for the instrument before use. 

OTHER FEATURES 

 

Degree of pollution 2 

 

Installation category II 

 

Use only in closed environments 

Summary of Contents for EasyPGX qPCR instrument 96

Page 1: ...tics srl requiring the use of EasyPGX qPCR instrument 96 in the respective user manual RT800 96 8057289090675 2017 06 Diatech Pharmacogenetics srl a Socio Unico Società soggetta all attività di direzione e coordinamento di Diatech srl con sede in Jesi Via Ignazio Silone 1 b 60035 Jesi AN Italy C F P Iva R Imprese di Ancona n 02483840423 Tel 39 0731 213243 Fax 39 0731 213239 info diatechpharmacogen...

Page 2: ...EasyPGX qPCR instrument 96 Cat Nr RT800 96 2 21 2017 06 This page was intentionally left blank ...

Page 3: ...ENVIRONMENT REQUIREMENTS 7 ELECTROSTATIC DISCHARGES 7 OTHER FEATURES 7 INSTALLATION 8 POSITION OF THE INSTRUMENT 8 UNPACKING OF THE INSTRUMENT 8 CLEANING THE HEATING BLOCK 8 CONNECT THE INSTRUMENT TO THE ELECTRIC POWER 8 CONNECT KEYBOARD AND MOUSE TO THE INSTRUMENT OPTIONAL 9 BACKGROUD CALIBRATION 9 DATE AND TIME SETTING 9 DIAGNOSTIC TEST 9 SOFTWARE INSTALLATION 10 ARIA DX SOFTWARE INSTALLATION 10...

Page 4: ...ION 15 TOUCH SCREEN CALIBRATION 15 INSTRUMENT MANTEINANCE 15 CLEANING THE INSTRUMENT 15 CLEANING THE OPTICAL MODULES 15 TECHNICAL SPECIFICATIONS 17 WARRANTY AND CLAUSE OF LIABILITY 18 WARRANTY 18 EXCLUSIONS 18 GUARANTEE LIMITATION AND REMEDIES 18 RESPONSIBILITY CLAUSE 18 APPENDIX A ELECTRIC AND ELECTRONIC WASTE DISPOSAL DIRECTIVE WEEE 19 OPERATOR NOTES 20 ...

Page 5: ...NOT PROVIDED USB pen drive USB 2 0 or higher if used also for execution of runs Personal computer Minimum requirements for running Aria Dx software Operating system Windows 7 Professional and Ultimate editions or Windows 10 Supported architectures o x86 32 bit supported on Windows 7 o x64 64 bit supported on Windows 7 and 10 Programs o Microsoft NET Framework 4 0 o Microsoft SQL Server 2012 only r...

Page 6: ...edical devices IVD only in countries that accept the user manual translated into the languages available on the web site www diatechpharmacogenetics com area riservata QUALITY ASSURANCE The EasyPGX centrifuge vortex 1 5 ml kit has been designed developed and validated in compliance with the Directive 98 79 EC on in vitro diagnostic IVD medical devices transposed in Italy in the D Lgs No 332 2000 a...

Page 7: ... to process inflammable explosive or reactive substances It is responsibility of the user to apply appropriate safety measures if potentially pathogenic material radioactive substances or other hazardous substances are used Do not immerse the instrument in any liquid DANGER OF BURNS Do not touch the thermoblock the inner part of the lid or the reaction tubes strips plates while the instrument is o...

Page 8: ...t stays supported on the box Lift the instrument and place it where it is expected The instrument weighs 23 kg To avoid damages to the instrument and people it is recommended to lift the instrument at least in two CLEANING THE HEATING BLOCK Whit the instrument cover opened clean the outer and inner surfaces of the heating block Lift the heat block door by pulling the handle forward and then upward...

Page 9: ... the Backgroud Calibration procedure fails contact Diatech Pharmacogenetics Technical Support email support diatechpharmacogenetics com phone 39 0731 213243 DATE AND TIME SETTING In the lower right corner of the display press the displayed date and time A menu window shows up press Change Date and Time In the window that appears set the correct date and time Click on the Help icon to see detailed ...

Page 10: ... diatechpharmacogenetics com to check for compatibility Otherwise simply install the Aria DX software on the computer that will be used for the analysis of the results To check the minimum requirements of the PC refer to the section Material Required but Not Provided The standard Windows Installer setup msi is available on request from Diatech Pharmacogenetics Technical Support email support diate...

Page 11: ...yPGX qPCR instrument 96 Cat Nr RT800 96 11 21 2017 06 The Application Mode window opens select AriaDX and press Next The Setup Type window opens Select Standard and press Next Do not change this selection ...

Page 12: ...17 06 The Destination Folder window opens Press Next to continue Press Install The wizard installs the AriaDX software in the folder shown When the installation is complete the InstallShield Wizard Completed window opens Click Finish to close the wizard ...

Page 13: ...armacogenetics srl It is recommended to use only pre set templates available on request contacting Diatech Pharmacogenetics Technical Support email support diatechpharmacogenetics com Import of template should only be performed the first time you use the specific kit associated with that template In subsequent runs it is sufficient to recall it see paragraph Execution of a run Save the template co...

Page 14: ...aved Experiment Select the last run file from the list Press Copy Select USB Disk on the left Press Paste Return to the home screen by pressing the Home button at the bottom left of the display and disconnect the USB stick ANALYSIS OF THE RESULTS For the analysis of the results obtained with the kit of the EasyPGX ready line refer to the user manuals that can be downloaded from the private area of...

Page 15: ...to optimize sensitivity On the Home screen press Settings In the Settings screen press Calibrate Touch The next screen shows how to use a marker Follow the instructions on the screen to complete the procedure INSTRUMENT MANTEINANCE CLEANING THE INSTRUMENT To clean EasyPGX qPCR instrument 96 follow the steps below Use a cloth dampened with water or isopropyl alcohol to clean the outside of the inst...

Page 16: ...yPGX qPCR instrument 96 detects an error it displays a special icon at the bottom of the touch display Press this icon In the popup menu that opens press Count X where X matches the number of error messages to read A table opens with the following information o Type message type error or warning Warning messages are used for errors that do not prevent the execution of experiments o ID Error Identi...

Page 17: ...Relative humidity 20 80 without condensation Maximum altitude 2000 m Excitation source 8 LEDs for each optical module Detection 8 photodiodes for each optical module Channels FAM HEX Power supply 100 240 VAC 50 60 Hz 1100 VA Thermal system Peltier block Positions 96 Temperature range of the thermal system 25 0 99 9 C Uniformity 0 2 5 C above 60 C 0 4 C 40 59 C Maximum heating speed 6 C s Maximum c...

Page 18: ...ND REMEDIES The warranty previously described is the unique and exclusive guarantee of Diatech Pharmacogenetics srl and the repair or replacement of defective parts is the sole and exclusive right of remedy No other warranties no express or implied warranties are provided The implied warranties of merchantability and eligibility for particular purposes are explicitly excluded extended to all categ...

Page 19: ...or environment The end user of the products must certify this with a specific Declaration of Decontamination which must be handed over to the carrier The End User is responsible for the deletion and or removal of any sensitive data contained in the product being returned Shipping costs will be charged to the Customer unless otherwise agreed with Diatech Pharmacogenetics Packaging conditions All pr...

Page 20: ...EasyPGX qPCR instrument 96 Cat Nr RT800 96 20 21 2017 06 OPERATOR NOTES ...

Page 21: ...Diatech Pharmacogenetics srl a Socio Unico via Ignazio Silone 1 b 60035 Jesi AN Italy Tel 39 0731 213243 Fax 39 0731 213239 info diatechpharmacogenetics com www diatechpharmacogenetics com ...

Reviews: