![Diamond SmartLyte Service Manual Download Page 39](http://html.mh-extra.com/html/diamond/smartlyte/smartlyte_service-manual_2493455039.webp)
4 Description of Modules
ECO 9085 SOP05-5038F Rev.05 Effective Date: 12/06/18 Page 39 of 81
4.1.8.
Power Supply Subassembly CN-A4019
The power supply subassembly is located on the rear panel and is comprised of the power
supply assembly and the main power receptacle. To remove the rear panel assembly,
ensure that the power cord has been disconnected. Remove the four corner screws to
expose the rear panel. Disconnect the ground wires and connector from the CPU board
and front case. The power supply module located on the rear panel assembly can now be
removed by removing the four screws securing the power supply mounting plate to the
rear panel. The main power receptacle can be replaced by removing the two screws on
the rear panel holding this assembly.
Fig. 4-9