1 Introduction
ECO 9099 SOP05-5090F Rev.00 Effective Date: 12/20/18 Page 36 of 58
1.3.1.12. RFID Board (CN-A4054)
The RFID board is attached to the CPU board at JP10. (
The RFID board can be accessed after removal of the CPU board as described in
This board must be connected all the time to the CPU board for proper functioning of the
instrument.
DO NOT unplug the RFID board while the CPU board is powered ON. Irreparable damage may
occur.
Figure 1-51 RFID Board
1.3.1.13. Valve Board (CN-A4018)
Valve board has multiple connections to devices on a single platform. The Aspiration Assembly,
Solenoid Valves, Peri-Pump Motor, all plug into the valve board. The lamp that lights the
measurement pathway is also located on this board. To access the board;
1. The Rear Panel and CPU with RFID boards, described in
2. Remove one screw with washer on the right side which holds the board in place. (Figure
4.36 below)
3. Then remove the seven electrical connectors, aspiration, solenoid valves, pump motor, to
slide this board out.
RFID
Antenna
RFID
Components
Connector to
CPU Board