Function description
OPERATION MANUAL
Chapter 3
Water Cooler
Errors and omissions excepted.
V1.1.0en/13.12.12
34
3.6.3
Setting the Auto-Start function (automatic mains failure)
Setting
Display
Description
0
The auto-start function is switched on
1
The auto-start function is switched off
PROCEDURE
Simultaneously press the
>Arrow keys<
and
for about 3 s. The display changes from the
temperature display to show the first menu item.
Press the
>Arrow key<
to display the menu point “C40”.
Press the
>SET key<
and hold it down.
Press the
>SET key<
and the
>Arrow keys<
and
simultaneously, the display toggles be-
tween “0” (auto-start function is switched on) to “I” (auto-start function is switched off). After
the desired setting is displayed, release the
>SET key<
.
Simultaneously press the
>Arrow keys<
and
for about 1 second. Or wait after releasing the
>SET key<
for a few seconds. The selected function is saved and the menu is exited. The display
returns to the temperature display.
Settings in the menu
point "C40" auto-start