background image

 

DVU, PPC, PPC2+, PPC3, and PPC BG0002 are trademarks, registered and otherwise, of 

DH Instruments, Inc. 

Teflon

® 

is a registered trademark of E.I. du Pont de Nemours and Company.

 

Document 560041a 

030211 

DH Instruments, Inc. 

4765 East Beautiful Lane 
Phoenix  AZ  85044-5318 
USA 

 

 

Tel 602.431.9100

Fax 602.431.9559

[email protected]

www.dhinstruments.com

 
 

 

PK-PPC-BG-DVU Dual Volume Unit 

and Interconnections Kit 

P/N 401699 

Installation Sheet

 

 

INTERCONNECTIONS KIT, P/N 401699, 
INCLUDES 

QTY PART 

NO 

FIGURE 

CALLOUT DESCRIPTION 

401697 

N/A 

Dual Volume Unit (DVU) 

101450-Z 

Tubing, PFA, 1/4 in. x 160 cm 

2 103279 

Adaptor, 1/4 in. tube x O-seal 
1/8 in. NPT M 

4 101477 

Adaptor, 1/4 in. tube x 1/4 in. 
NPT M 

2 101985 

4  Union, 

4P 

101487 

Tee, 1/4 in. tube 

2 101482 

Adaptor, 1/4 in. tube x 1/4 in. 
NPT F 

 

TO INSTALL THE DUAL VOLUME UNIT (DVU) 

  Identify a location for the DVU™ that minimizes 

the length of tubing between the DVU connections 
and the PPC BG0002™ TEST(+) and TEST(-) ports.  
The PPC BG0002 and the DVU are designed to 
stack one on top of the other.  The DVU may be 
rack mounted using the rack mount kit (P/N 400440).  
Recommended configurations are for the PPC and 
DVU to be stacked directly one on top of the other 
or with their rear panels face to face. 

TO CONNECT THE DVU AND PPC BG0002 

 

 

You will be making and connecting two identical 
assemblies, one for the TEST(+) port and one for 
the TEST(-) port. 

 

  Install a union (4) into a 1/4 in. tube x 1/4 in. NPT 

M adaptor (3) (do this twice).  Screw the 1/4 in. 
NPT M end of the adaptors into the two ports of 
the DVU using Teflon

®

 tape as a sealer. 

 

 

Figure 1  

Interconnections Kit 

 

  Install a 1/4 in. tube tee (5) onto each of the 

union/adaptor assemblies that were just installed 
into the DVU ports in Step  .  Orient the tee 
considering the connection to the PPC BG0002 
(schematic above is for PPC stacked on top of  
the DVU). 

  Install the appropriate adaptors onto the PPC BG0002 

TEST(+) and TEST(-) ports:  

For PPC3™ 

Install the 1/4 in. tube x O-seal  
1/8 in. NPT M (2). 

For PPC2+™

  Install the 1/4 in. tube x 1/4 in.  

NPT M adaptor (3) using Teflon 
tape as a sealer. 

  Cut lengths of PFA tubing as needed to connect 

the 1/4 in. tube tee to the adaptors on the PPC 
TEST ports.  Take care that the cut is at a right 
angle to the tube and only cut the length needed. 

  Optional for a connection “For DUT” of 1/4 in. NPT M 

rather than 1/4 in. tube: 

•  Cut a small length of PFA tubing to use as  

a union between the 1/4 in. tube tee (5) and  
the 1/4 in. tube x 1/4 in. NPT F adaptor (6) 
(see schematic above). 

 

Reviews: